Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 87868

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
April 07, 2021
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
RAYSEARCH LABORATORIES AB

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

RayStation standalone software radiation therapy treatment planning system. Model Number: 8A, 8B, 9A, 9B, 10A, 10B including some service packs

Z-1598-2021
Recall number
Z-1598-2021
Initiated
April 07, 2021
Classification
Class II
Status
Terminated
Recalling firm
RAYSEARCH LABORATORIES AB
Quantity
N/A

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
When using more than one plan in a treatment course, the Total dose display in Dose tracking may be incorrect. There may be a mismatch between the fractions that appear as selected in the Treatment course list in Dose tracking and the fractions that are included in the dose summation. The issue only affects RayTreat users and only the Total dose option. Accumulation and other options in Dose tracking are correct.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

When using more than one plan in a treatment course, the Total dose display in Dose tracking may be incorrect. There may be a mismatch between the fractions that appear as selected in the Treatment course list in Dose tracking and the fractions that are included in the dose summation. The issue only affects RayTreat users and only the Total dose option. Accumulation and other options in Dose tracking are correct.

Code information

Software version number: 8.0.0.6 , 8.0.1.10 , 8.1.0.47 , 8.1.1.8 , 8.1.2.5 , 9.0.0.113 , 9.1.0.933 , 9.2.0.483 , 10.0.0.1154 , 10.0.1.52 , 10.1.0.613

Distribution pattern

Domestic: TN