Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 87999

5 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
April 08, 2021
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Spes Medica

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

5 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 5

Dragonfly Laryngeal Surface Electrode, ET Tube Size (mm): 7.5-10.0, Channel 1, Reference Numbers LSE4246MCA075 and LSE500M. for continuous EMG monitoring of the larynx during surgical procedures.

Z-1967-2021
Recall number
Z-1967-2021
Initiated
April 08, 2021
Classification
Class II
Status
Terminated
Recalling firm
Spes Medica
Quantity
20 boxes

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The supplier informed them of possible deviations of the parameters/processes defined for ethylene oxide sterilization.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Material/Component Contamination

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The supplier informed them of possible deviations of the parameters/processes defined for ethylene oxide sterilization.

Code information

Lot LS20G3274I

Distribution pattern

Distribution in the following US states: CA, MA, and SC.

device · product 2 of 5

SNS Disposable Urethral Catheter Electrode, 10 Fr, Gold Recording/Stimulating Contact printed on polyurethane film adhesive with attached 2 meter lead-wire with touch-proof connector, Reference Number UE006

Z-1968-2021
Recall number
Z-1968-2021
Initiated
April 08, 2021
Classification
Class II
Status
Terminated
Recalling firm
Spes Medica
Quantity
34 boxes

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The supplier informed them of possible deviations of the parameters/processes defined for ethylene oxide sterilization.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Material/Component Contamination

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The supplier informed them of possible deviations of the parameters/processes defined for ethylene oxide sterilization.

Code information

Lot UR20F2702I

Distribution pattern

Distribution in the following US states: CA, MA, and SC.

device · product 3 of 5

Dragonfly Laryngeal Surface Electrode, ET Tube Size (mm): 6.0-7.5, Channel 2, Reference Number LSE3463DCA200 and LSE500DCS-5. for continuous EMG monitoring of the larynx during surgical procedures.

Z-1969-2021
Recall number
Z-1969-2021
Initiated
April 08, 2021
Classification
Class II
Status
Terminated
Recalling firm
Spes Medica
Quantity
8 boxes

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The supplier informed them of possible deviations of the parameters/processes defined for ethylene oxide sterilization.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Material/Component Contamination

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The supplier informed them of possible deviations of the parameters/processes defined for ethylene oxide sterilization.

Code information

Lot LS20E2538I and 110320C

Distribution pattern

Distribution in the following US states: CA, MA, and SC.

device · product 4 of 5

SNS Disposable Urethral Catheter Electrode, 14 Fr, Gold Recording/Stimulating Contact printed on polyurethane film adhesive with attached 2 meter lead-wire with touch-proof connector, Reference Number UE002

Z-1970-2021
Recall number
Z-1970-2021
Initiated
April 08, 2021
Classification
Class II
Status
Terminated
Recalling firm
Spes Medica
Quantity
30 boxes

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The supplier informed them of possible deviations of the parameters/processes defined for ethylene oxide sterilization.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Material/Component Contamination

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The supplier informed them of possible deviations of the parameters/processes defined for ethylene oxide sterilization.

Code information

Lot UR20F2699I

Distribution pattern

Distribution in the following US states: CA, MA, and SC.

device · product 5 of 5

Tripolar Hooked Stimulation Probes, L=20mm, Reference Number SI3H0020S2526D. Used during intraoperative neurological monitoring.

Z-1971-2021
Recall number
Z-1971-2021
Initiated
April 08, 2021
Classification
Class II
Status
Terminated
Recalling firm
Spes Medica
Quantity
3 boxes

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The supplier informed them of possible deviations of the parameters/processes defined for ethylene oxide sterilization.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Material/Component Contamination

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The supplier informed them of possible deviations of the parameters/processes defined for ethylene oxide sterilization.

Code information

Lot SI20F2956I

Distribution pattern

Distribution in the following US states: CA, MA, and SC.