device · product 1 of 1
The ADVIA Chemistry XPT Chemistry System, Siemens Material Number (SMN) 10723034, is an automated, clinical chemistry analyzer that runs tests on serum, plasma, urine, or cerebral spinal fluid in random access and batch modes. Tests performed using this system are intended for in vitro diagnostic use.
- Recall number
- Z-2014-2021
- Initiated
- May 18, 2021
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 842 units
App-derived interpretation
Software Versions (V1.4 SMN 11314625 and V1.4 China SMN 11316885) may
Official device-enrichment evidence · Sourced
Software design
Inspect official wording and provenance
Reason for recall
ADVIA Chemistry System and Software Versions (V1.4 SMN 11314625 and V1.4 China SMN 11316885) may not be aligned with customer expectations for the current default mixer failure detection setting of 2; an erroneously elevated or depressed patient result to be reported without alert on the test result
Code information
Software version: 1.4 or lower UDI - 00630414595467
Distribution pattern
Worldwide distribution - US Nationwide distribution in the states of CA, CO, CT, FL, MA, MD, MI, MN, NY, OH, PA, TX and the countries of Australia, Bahrain, Brazil, Canada, Chile, Croatia, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, Italy, Malaysia, Netherlands, New Zealand, Norway, P.R. China, Portugal, Republic Korea, Saudi Arabia, Slovakia, Spain, Sweden, Taiwan, Thailand, Turkey, United Kingdom, Vietnam.