Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 88187

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 17, 2021
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Gentherm Medical, LLC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Hemotherm CE Dual Reservoir Cooler/Heater, Model 400CE, 115V Model, part no. 86022. Used to cool or warm blood during cardiopulmonary bypass procedures

Z-2092-2021
Recall number
Z-2092-2021
Initiated
June 17, 2021
Classification
Class II
Status
Terminated
Recalling firm
Gentherm Medical, LLC
Quantity
1448 devices (1010 US; 438 OUS)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
There is a potential risk of device contamination and patient infection associated with the device because of the potential for organisms (including Nontuberculous mycobacteria NTM) to grow in the water systems of any heater-cooler device, and contaminated water from any heater-cooler device has the potential to aerosolize into the operating room during surgery which could lead to patient infection.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

There is a potential risk of device contamination and patient infection associated with the device because of the potential for organisms (including Nontuberculous mycobacteria NTM) to grow in the water systems of any heater-cooler device, and contaminated water from any heater-cooler device has the potential to aerosolize into the operating room during surgery which could lead to patient infection.

Code information

All units in service life Serial Numbers: 092-10011CE tot 212-11499CE

Distribution pattern

Worldwide distribution. US nationwide, including Puerto Rico; Argentina, Bangladesh, Chile, Colombia, France, Guatemala, Hungary, India, Indonesia, Israel, Italy, Nepal, Nigeria, Pakistan, Philippines, Romania, Saudi Arabia, Singapore, South Africa, South Korea, Switzerland, Taiwan, Thailand, Turkey, UAE