openFDA Device Enforcement
Report-date coverage
June 20, 2012–July 08, 2026
Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
openFDA Device Recall
event_date_initiated coverage
June 01, 1997–June 30, 2026
Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.
Product dates, classifications, firms, and source wording remain attached to their individual rows.
Opening this dossier never hides products that did not match a prior timeline filter.
device · product 1 of 1
Infusomat Space Volumetric Pump Administration Set, Product Code 490100
There is a potential for the Anti-free flow clip of the administration set to be inverted which may cause unintended free flow if the user fails to follow on screen prompt to close the roller clamp prior to set removal from the pump.
These labels are deterministic app interpretations, not FDA categories.
There is a potential for the Anti-free flow clip of the administration set to be inverted which may cause unintended free flow if the user fails to follow on screen prompt to close the roller clamp prior to set removal from the pump.
Code information
Lot Number: 0061761278
Distribution pattern
Affected products were distributed to the following US states: CA, MI, OR, and UT.