Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 88266

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 16, 2021
Product types
Device
Classifications
Class I
Statuses
Terminated
Recalling firm wording
Cardinal Health 200, LLC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Argyle UVC Insertion Tray containing Safety Scalpel N11 (vendor part number 73-1811). Intended for use to insert an Umbilical Venous Catheter (UVC) into the umbilical artery or vein of neonate. Product Code: 43201

Z-2244-2021
Recall number
Z-2244-2021
Initiated
June 16, 2021
Classification
Class I
Status
Terminated
Recalling firm
Cardinal Health 200, LLC
Quantity
7197 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The current Argyle UVC Insertion Tray does not include a specific IFU to operate the Safety Scalpel N11. The Safety Scalpel N11 features a permanent locking feature that will not allow the clinician to unlock the scalpel once the safety shield is closed and pushed forward into a locked position may result in a delay treatment/therapy, which could ultimately result in death.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Labeling Change Control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The current Argyle UVC Insertion Tray does not include a specific IFU to operate the Safety Scalpel N11. The Safety Scalpel N11 features a permanent locking feature that will not allow the clinician to unlock the scalpel once the safety shield is closed and pushed forward into a locked position may result in a delay treatment/therapy, which could ultimately result in death.

Code information

All lot codes

Distribution pattern

US Nationwide distribution.