Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 88333

4 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
July 26, 2021
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Philips Respironics, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

4 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 4

Trilogy Evo, Product numbers BL2110X15B, CA2110X12B, DE2110X13B, DS2110X11B, EE2110X15B, ES2110X15B, EU2110X15B, EU2110X19, FR2110X14B, GB2110X15B, IA2110X15B, IN2110X15B, IN2110X19, IT2110X21B, KR2110X15B, LA2110X15B, LD2110X23B, ND2110X15B, RDE2110X13B, UDS2110X11B

Z-2287-2021
Recall number
Z-2287-2021
Initiated
July 26, 2021
Classification
Class II
Status
Ongoing
Recalling firm
Philips Respironics, Inc.
Quantity
18588 (US); 5662 (OUS)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Two software issues have been identified related to pressure increase in the device: The first issue is described as "Infant/Pediatric EFS Calibration Pressure Increase" and the second issue is described as "Pressure Drift (Continuous Usage).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Labeling design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Two software issues have been identified related to pressure increase in the device: The first issue is described as "Infant/Pediatric EFS Calibration Pressure Increase" and the second issue is described as "Pressure Drift (Continuous Usage).

Code information

Software version numbers 1.02.01.00, 1.03.05.00, 1.03.07.00, 1.04.02.00, 1.04.06, 1.05.01, and 1.06.02

Distribution pattern

Worldwide distribution - US nationwide distribution.

device · product 2 of 4

Trilogy Evo O2, Product numbers DE2100X13B, DS2100X11B, EE2100X15B, ES2100X15B, EU2100X15B, FR2100X14B, FX2100X15B, IA2100X15B, IN2100X15B, IN2100X19, IT2100X21B, JP2100X16B, LA2100X15B, ND2100X15B, RDE2100X13B, SP2100X26B

Z-2288-2021
Recall number
Z-2288-2021
Initiated
July 26, 2021
Classification
Class II
Status
Ongoing
Recalling firm
Philips Respironics, Inc.
Quantity
101 (FEMA Stockpile Only); 21077 (OUS)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Two software issues have been identified related to pressure increase in the device: The first issue is described as "Infant/Pediatric EFS Calibration Pressure Increase" and the second issue is described as "Pressure Drift (Continuous Usage).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Labeling design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Two software issues have been identified related to pressure increase in the device: The first issue is described as "Infant/Pediatric EFS Calibration Pressure Increase" and the second issue is described as "Pressure Drift (Continuous Usage).

Code information

Software version numbers 1.02.01.00, 1.03.05.00, 1.03.07.00, 1.04.02.00, 1.04.06, 1.05.01, and 1.06.02

Distribution pattern

Worldwide distribution - US nationwide distribution.

device · product 3 of 4

Trilogy EV300, Product numbers CA2200X12B, DS2200X11B, FX2200X15B, IN2200X15B

Z-2289-2021
Recall number
Z-2289-2021
Initiated
July 26, 2021
Classification
Class II
Status
Ongoing
Recalling firm
Philips Respironics, Inc.
Quantity
18936 (US); 4659 (OUS)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Two software issues have been identified related to pressure increase in the device: The first issue is described as "Infant/Pediatric EFS Calibration Pressure Increase" and the second issue is described as "Pressure Drift (Continuous Usage).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Labeling design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Two software issues have been identified related to pressure increase in the device: The first issue is described as "Infant/Pediatric EFS Calibration Pressure Increase" and the second issue is described as "Pressure Drift (Continuous Usage).

Code information

Software version numbers 1.02.01.00, 1.03.05.00, 1.03.07.00, 1.04.02.00, 1.04.06, 1.05.01, and 1.06.02

Distribution pattern

Worldwide distribution - US nationwide distribution.

device · product 4 of 4

Trilogy Evo Universal, Product number DS2000X11B

Z-2290-2021
Recall number
Z-2290-2021
Initiated
July 26, 2021
Classification
Class II
Status
Ongoing
Recalling firm
Philips Respironics, Inc.
Quantity
10,000 (FEMA Stockpile Only)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Two software issues have been identified related to pressure increase in the device: The first issue is described as "Infant/Pediatric EFS Calibration Pressure Increase" and the second issue is described as "Pressure Drift (Continuous Usage).

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Labeling design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Two software issues have been identified related to pressure increase in the device: The first issue is described as "Infant/Pediatric EFS Calibration Pressure Increase" and the second issue is described as "Pressure Drift (Continuous Usage).

Code information

Software version numbers 1.02.01.00, 1.03.05.00, 1.03.07.00, 1.04.02.00, 1.04.06, 1.05.01, and 1.06.02

Distribution pattern

Worldwide distribution - US nationwide distribution.