device · product 1 of 1
Terumo Advanced Perfusion System 1 Electronic Patient Gas System Catalog #: 801188
intermittent failure causes a blender initialization fault upon start-up, or during an EPGS calibration attempt following the 15-minute warm-up period. The control sliders on the Central Control Monitor (CCM) are disabled and the gas system can only be controlled by the local control knobs
These labels are deterministic app interpretations, not FDA categories.
intermittent failure causes a blender initialization fault upon start-up, or during an EPGS calibration attempt following the 15-minute warm-up period. The control sliders on the Central Control Monitor (CCM) are disabled and the gas system can only be controlled by the local control knobs
Code information
S/N: 01751 UDI: 00886799000588
Distribution pattern
International distribution in the country of Panama.