Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 88351

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
July 22, 2021
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Baxter Healthcare Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

Baxter PrismaFlex, REF numbers: a) 106696 (M 60 SET) b) 106697 (M 100 SET) Expansion: c) 107144 (TPE2000 SET) d) 115307 (M 150 SET) e) 115309 (ST 100 SET) f) 955467 (M 150 SET) g) 955468 (ST 100 SET) h) 955595 (M 100 SET) i) 955596 (ST 150 SET) Used with Prismaflex or PrisMax control units for continuous fluid management and renal replacement therapies.

Z-2378-2021
Recall number
Z-2378-2021
Initiated
July 22, 2021
Classification
Class II
Status
Terminated
Quantity
64275 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Products labeled as sterile were distributed, but may not have been sterilized. " Baxter confirmed Steril Milano observed nonconformances that were not reported to Baxter. Data was manipulated to appear to be conforming

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Vendor change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Products labeled as sterile were distributed, but may not have been sterilized. " Baxter confirmed Steril Milano observed nonconformances that were not reported to Baxter. Data was manipulated to appear to be conforming

Code information

a) 106696, Lot Numbers: 20B2329M b) 106697, Lot Numbers: 20B2323M, 20B2328M (expanded to include 20B2326M) Expansion: c) 107144, Lot Number 20B2325M d) 115307, Lot Number 20B2321M e) 115309, Lot Number 20B2310M and 20F0201 f) 955467, Lot Numbers 20A2604M, 20A2606M, 20A2607M, 20A2608M, 20B2331M, 20B2332M, 20B2333M, 20B2334M, 20B2335M, 20B2336M, 20B2347M, 20A2605M g) 955468, Lot Numbers 20A2602M, 20A2603M, 20B2308M, 20B2311M, 20B2312M, and 20B2324M h) 955595, Lot Numbers 20A2609M, 20A2610M, 20B2337M, 20B2338M, 20B2339M, 20B2340M, 20B2341M, 20B2342M, 20B2343M, 20B2344M, 20B2345M, and 20B2346M i) 955596, Lot Numbers 20A2601M, 20B2302M, and 20B2309M

Distribution pattern

US Nationwide distribution.

device · product 2 of 2

Baxter prismaflex, HF 1000 PAES membrane, REF 107140 Used with Prismaflex or PrisMax control units for continuous fluid management and renal replacement therapies.

Z-2379-2021
Recall number
Z-2379-2021
Initiated
July 22, 2021
Classification
Class II
Status
Terminated
Quantity
1976 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Products labeled as sterile were distributed, but may not have been sterilized. Baxter confirmed Steril Milano observed nonconformances that were not reported to Baxter Data was manipulated to appear to be conforming

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Vendor change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Products labeled as sterile were distributed, but may not have been sterilized. Baxter confirmed Steril Milano observed nonconformances that were not reported to Baxter Data was manipulated to appear to be conforming

Code information

Lot Numbers: 20B2330M

Distribution pattern

US Nationwide distribution.