Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 88412

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 30, 2021
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Siemens Medical Solutions USA, Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Artis Models with SW version VD12 listed below: Artis zee floor 10094135 Artis zee ceiling 10094137 Artis zee MP 10094139 Artis Q floor 10848280 Artis Q ceiling 10848281 Artis Q biplane 10848282 Artis Q.zen floor 10848353 Artis Q.zen ceiling 10848354 Artis Q.zen biplane 10848355 Angiography systems developed for single and biplane diagnostic imaging

Z-2404-2021
Recall number
Z-2404-2021
Initiated
June 30, 2021
Classification
Class II
Status
Terminated
Quantity
70 US (172 WW)

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
software error

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Due to a software error, the IAS (Image Acquisition System) may sporadically fail during startup/ restart, and result in delay of clinical treatment

Code information

Units with SW version VD12 Serial Numbers: 104005 104007 104012 104013 105500 105501 109645 110001 110005 110006 110007 110014 110016 110017 110018 110022 110023 110024 110025 110026 110027 110031 110032 110035 110036 110037 110039 110040 110041 110042 110043 110044 110046 110048 110051 111133 111500 111501 111502 111503 111504 111505 111506 111507 111508 111510 122001 122002 122005 122007 123501 123502 139012 150002 150003 150005 150009 150011 159010 159015 159021 159024 159025 159026 159028 159030 159032 159034 159037 159038

Distribution pattern

US Nationwide distribution in the states of AR, AZ, CA, CO, FL, GA, IA, ID, IL, IN, KY, MA, MD, MI, MN, MO, ND, NE, NH, NJ, NV, NY, OH, PA, TN, TX, UT, VA, WA, WI.