device · product 1 of 1
Proteus 235- designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation
- Recall number
- Z-2434-2021
- Initiated
- June 14, 2021
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Ion Beam Applications S.A.
- Quantity
- 27 worldwide, 11 in U.S.
App-derived interpretation
Users may be misled by the popup message displayed by the Proton Therapy System when the Oncology Information System appears disconnected, may cause mistreatment
Official device-enrichment evidence · Sourced
Device Design
Inspect official wording and provenance
Reason for recall
Users may be misled by the popup message displayed by the Proton Therapy System when the Oncology Information System appears disconnected, may cause mistreatment
Code information
PAT.006 (US), PAT.107 (EU), PAT.108 (US), PAT.110 (US), PAT.112 (US), PAT.113 (US), PAT.114 (EU), PAT.115 (EU), SAT.116 (US), SAT.118 (RU), SAT.119 (US), SAT.120 (EU), SAT.122 (EU), SAT.123 (US), SAT.125 (IN), SAT.126 (US), SAT.132 (EU), SBF.101 (EU), SBF.103 (JP), SBF.104 (JP), SBF.105 (US), SBF.107 (EU), SBF.109 (UK), SBF.110 (UK), SBF.113 (US), SBF.115 (UK), SBF.117 (EU).
Distribution pattern
Worldwide distribution - US Nationwide distribution in the states of FL, OK, VA, IL, NJ, TN, WA, LA, TX, MI and the countries of Germany, Italy, Russia, Sweden, India, Japan, France, England, Spain.