Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 88541

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
August 06, 2021
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Heartware, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Medtronic HVAD Pump Implant Kits

Z-2478-2021
Recall number
Z-2478-2021
Initiated
August 06, 2021
Classification
Class II
Status
Ongoing
Recalling firm
Heartware, Inc.
Quantity
11851 (potentially active, US); 1995 (active, US)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
There is more force required to slide the driveline cover (boot cover) away from the connector after completion of a Driveline Strain Relief Repair than over an unrepaired strain relief. This may lead to difficulty in pulling back the driveline cover when attempting to access the Driveline connector during a controller exchange. Additionally, sliding the driveline cover over a strain relief repair may compromise the repair.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Device Design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

There is more force required to slide the driveline cover (boot cover) away from the connector after completion of a Driveline Strain Relief Repair than over an unrepaired strain relief. This may lead to difficulty in pulling back the driveline cover when attempting to access the Driveline connector during a controller exchange. Additionally, sliding the driveline cover over a strain relief repair may compromise the repair.

Code information

Product Numbers: 1100, 1101, 1102, 1103, 1104, 1104JP, 1205, and MCS1705PU - all serial numbers

Distribution pattern

Distribution throughout the US. Also distribution in ARGENTINA Armenia, Australia, Austria, BELGIUM, BRAZIL, CANADA, CHILE, Colombia, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA, FINLAND, FRANCE, Germany, GREECE HONG KONG, HUNGARY, INDIA, ISRAEL, Italy, JAPAN, Kazakhstan, Korea, KUWAIT, Latvia, LEBANON, LITHUANIA, LUXEMBOURG, Macedonia, MALAYSIA, NETHERLANDS, NEW ZEALAND, NORWAY, Poland, ROMANIA, SAUDI ARABIA, Serbia, Singapore, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, Taiwan, TURKEY, Ukraine, United Arab Emirates, United Kingdom, VIETNAM