Timeline bucket
August 18, 2021
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Siemens Medical Solutions USA, Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Computed tomography x-ray systems with software syngo.CT VA20A_SP4a, VA20A_SP5, VA30A_SP2, VA30A_SP2a, VA30A_SP3,VA30A_FP2 in Somatom systems: SOMATOM go.Up - Model 11061620 SOMATOM go.Up - Model 11061628 SOMATOM go.All - Model 11061630 SOMATOM go.Top - Model 11061640 SOMATOM X.cite - Model 11330001 Computed tomography systems intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data.

Z-2500-2021
Recall number
Z-2500-2021
Initiated
August 18, 2021
Classification
Class II
Status
Ongoing
Quantity
31 units US

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
Software versions may result in sporadic problems causing scanning workflow interruptions

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Software versions may result in sporadic problems causing scanning workflow interruptions and unexpected user notifications. Sporadic software errors may also occur during interventional workflows, resulting in delay in diagnosis or scan aborts with the necessity for patient rescan may occur

Code information

US Serial Numbers: 111307, 111526, 119328, 108158, 123104, 111219, 123054, 117633, 111566, 111427, 123059, 117558, 117533, 119317, 117647, 108088, 123060, 123118, 123055, 111518, 123113, 108108, 123129, 123122, 123069, 111507, 111560, 119142, 108111, 108062, 119376. Update CT016/21/S is a remote update for the following products with software versions syngo CT VA30A SP2 or syngo CT VA30A SP2a or syngo CT VA30A SP3: SOMATOM go.Now (Model #11061610, #11061612, #11061613, #11061618), SOMATOM go.Up (Model #11061620, #11061622, #11061623, #11061628), SOMATOM go.All (Model #11061630, #11061632, #11061638), SOMATOM go.Top (Model #11061640, #11061642, #11061648), SOMATOM go.Sim (Model #11061660, #11061668), SOMATOM go.Open Pro (Model #11061670, #11061678). Update CT018/21/S is a remote update for the following products with software versions syngo.CT VA20A_SP4a or syngo.CT VA20A_SP5: SOMATOM go.Now (Model #11061610, #11061612, #11061613, #11061618), SOMATOM go.Up (Model #11061620, #11061622, #11061623, #11061628), SOMATOM go.All (Model #11061630, #11061632, #11061638), SOMATOM go.Top (Model #11061640, #11061642, #11061648) without dual energy post-processing. Update CT022/21/S is an onsite update for the following product with software versions syngo.CT VA20A_SP4a or syngo.CT VA20A_SP5: SOMATOM go.Top (Model #11061640, #11061642, #11061648) without dual energy post-processing. Update CT025/21/S is an onsite update for the following product with software versions syngo.CT VA30A_SP2, VA30A_SP2a, syngo CT VA30A_SP3, or syngo CT VA30A_FP2 : SOMATOM X.cite (Model #11330001)

Distribution pattern

US Nationwide distribution.