Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 88676

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
September 14, 2021
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Elekta Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Elekta Monaco - Product Usage: used to make treatment plans for patients with prescriptions for external beam radiation therapy.

Z-2563-2021
Recall number
Z-2563-2021
Initiated
September 14, 2021
Classification
Class II
Status
Ongoing
Recalling firm
Elekta Inc
Quantity
214 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Contour changes can be saved on an unintended image set. In addition, these contour edits do not cause the dose to be frozen on the plans associated with that image set.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Contour changes can be saved on an unintended image set. In addition, these contour edits do not cause the dose to be frozen on the plans associated with that image set.

Code information

Software Builds: 5.40.00, 5.40.01, 5.40.02, 5.51.10; UDI GTIN: (01)00858164002190(10) 5.40.00, (01)00858164002190(10) 5.40.01, (01)00858164002190(10) 5.40.02, (01)00858164002268(10) 5.51.10

Distribution pattern

Worldwide distribution - US Nationwide distribution in the states of PA, MI, WI, NY, TN, NJ, IA, TX and the countries of United Kingdom, Turkey, Switzerland, Sweden, Spain, Netherlands, Korea, Thailand, Japan, Italy, Hong Kong, Germany, France, Denmark, China, Canada, Belgium, Bahrain, Australia.