Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 88681

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
September 17, 2021
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Elekta, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Elekta Medical Linear Accelerator (EMLA), configured as follows: a. Elekta Precise Digital Accelerator; b. Elekta Harmony; c. Elekta Synergy; d. Elekta Infinity; e. Versa HD

Z-0116-2022
Recall number
Z-0116-2022
Initiated
September 17, 2021
Classification
Class II
Status
Terminated
Recalling firm
Elekta, Inc.
Quantity
4,405 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
There is a risk that the Diode D1 in the Modulator can fail in a destructive manner when removed from the machine, and that the protective tube surrounding the diode may fail to contain the ejected matter.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

There is a risk that the Diode D1 in the Modulator can fail in a destructive manner when removed from the machine, and that the protective tube surrounding the diode may fail to contain the ejected matter.

Code information

a. Elekta Precise Digital Accelerator - UDI GTIN Number: 5060191071017, Model Numbers: XRT 0601 (NFB), XRT 0611 (FB), All Serial Numbers; b. Elekta Harmony - UDI GTIN Number: 05060191071581, Model Numbers: XRT 2201 (NFB), All Serial Numbers; c. Elekta Synergy - UDI GTIN Number: 05060191071550, Model Numbers: XRT 0621 (NFB), XRT 0631 (FB), All Serial Numbers; d. Elekta Infinity - UDI GTIN Number: 05060191071543, 05060191071512, Model Numbers: XRT 0511 (NFB & MLCi2), XRT 0521 (FB & MLCi2), XRT 0531 (NFB & Agility), XRT 0541 (FB & Agility), All Serial Numbers; e. Versa HD - UDI GTIN Number: 05060191071574, 05060191071529, Model Numbers: XRT 2121 (NFB), XRT 2131 (FB), All Serial Numbers

Distribution pattern

Distributed nationwide to FL, NC, OH, CA, IL, MD, OR, AL, NY, PA, MA, AK, GA, OR, KY, VA, TX, SD, AZ, NJ, MI, WI, ND, MO, WA, OK, PR, NM, NC, ID, CT, DE, LA, MT, TN, CO, UT, MN, IN, RI, IA, DC, ND, AZ, HI, KS, NE, NH, AR, WY, WV, MS, VT and internationally to Albania, Antigua/Barbuda, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegowina, Botswana, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Finland, France, Gabon, Georgia, Germany, Ghana, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Kosovo, Kyrgystan, Lao People's Democratic Republic, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Mozambique, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Pakistan, Palestinian Territory, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Republic of Serbia, Romania, Russian Federation, Rwanda, Saudi Arabia, Senegal, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Suriname, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Tunisia, Togo, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Vietnam, Yemen, Zimbabwe.