Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 88719

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
September 21, 2021
Product types
Device
Classifications
Class I
Statuses
Ongoing
Recalling firm wording
Datascope Corp.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Cardiosave Li-Ion Battery Pack, Part No. 0146-00-0097. Component of the Cardiosave Hybrid Intra-Aortic Balloon Pump (IABP) with the below Model Numbers. 1. Cardiosave Hybrid. Model Numbers: 0998-00-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-00-0800-35, 0998-00-0800-36, 0998-00-0800-45, 0998-00-0800-52, 0998-00-0800-53, 0998-00-0800-55, 0998-00-0800-65, 0998-UC-0800-31, 0998-UC-0800-33, 0998-UC-0800-52, 0998-UC-0800-53, 0998-UC-0800-55. 2. Cardiosave Rescue. Model Number: 0998-00-0800-83.

Z-0091-2022
Recall number
Z-0091-2022
Initiated
September 21, 2021
Classification
Class I
Status
Ongoing
Recalling firm
Datascope Corp.
Quantity
131 batteries, 11,517 pumps

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
A Cardiosave Lithium-Ion Battery Pack used during protocol testing failed to meet the minimum runtime requirement per Getinge internal Product Specification. These nonconforming batteries were inadvertently released to customers.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

A Cardiosave Lithium-Ion Battery Pack used during protocol testing failed to meet the minimum runtime requirement per Getinge internal Product Specification. These nonconforming batteries were inadvertently released to customers.

Code information

Battery Pack Serial Numbers: 192224349IP, 191908720IP, 202369811IP, 202791944IP, 191848115IP, 192099841IP, 192103741IP, 192104141IP, 171218521PE, 181642339PE, 202321308IP, 191912520IP, 191874716IP, 191922620IP, 181642639PE, 192185446IP, 171277935PE, 192190247IP, 202400219IP, 202323808IP, 191910720IP, 181661142PE, 192245250IP, 202599332IP, 181645841PE, 181526530PE, 202403719IP, 181705244PE, 192104541IP, 191787312IP, 181646041PE, 191955422IP, 202389919IP, 191822513IP, 191853415IP, 192004724IP, 192196447IP, 192155645IP, 202330609IP, 171325749PE, 192018828IP, 191778911IP, 192187947IP, 192196247IP, 181609237PE, 192098341IP, 191972823IP, 191860615IP, 202719740IP, 202334109IP, 202651035IP, 191852615IP, 192173146IP, 202386519IP, 202398619IP, 202416720IP, 202599432IP, 191956322IP, 192205847IP, 192220349IP, 192081138IP, 202617233IP, 181659242PE, 202266705IP, 171255732PE, 192179046IP, 192179246IP, 202280605IP, 181673143PE, 192083838IP, 192106442IP, 192216649IP, 191956122IP, 181721730IP, 191943120IP, 191971023IP, 181405120PE, 181474326PE, 181440222PE, 192225649IP, 191969623IP, 192220749IP, 181706344PE, 181694644PE, 191968223IP, 202593632IP, 191863015IP, 171264133PE, 191872616IP, 181487828PE, 171337749PE, 181477426PE, 181748032IP, 191790412IP, 191902319IP, 192195547IP, 192200947IP, 191968723IP, 202552831IP, 191806513IP, 202401419IP, 191904319IP, 181492328PE, 202388319IP, 202482724IP, 192241550IP, 171339849PE, 191831214IP, 192084038IP, 181694244PE, 192210249IP, 202725540IP, 171345550PE, 171348550PE, 191915920IP, 202371511IP, 202723440IP, 202367411IP, 202279305IP, 181513630PE, 181703744PE, 202389819IP, 192175646IP, 192107842IP, 202394719IP, 202401619IP, 192204947IP, 202654636IP, 192196647IP, 192200647IP, 202782744IP. 1. Cardiosave Hybrid. Model Number (UDI Number): 0998-00-0800-31 (10607567109053), 0998-00-0800-32 (10607567111117), 0998-00-0800-33 (10607567109008), 0998-00-0800-35 (10607567109107), 0998-00-0800-36 (10607567114187), 0998-00-0800-45 (10607567108421), 0998-00-0800-52 (10607567108438), 0998-00-0800-53 (10607567108391), 0998-00-0800-55 (10607567108414), 0998-00-0800-65 (10607567113432), 0998-UC-0800-31 (10607567109053), 0998-UC-0800-33 (10607567109008), 0998-UC-0800-52 (10607567108438), 0998-UC-0800-53 (10607567108391), 0998-UC-0800-55 (10607567108414). 2. Cardiosave Rescue. Model Number: 0998-00-0800-83 (10607567108407).

Distribution pattern

Worldwide distribution. US nationwide, Australia, Brazil, Bahrain, Belgium, Canada, China, Czech Republic, Egypt, France, Germany, India, Ireland, Italy, Japan, Kuwait, Netherlands, Norway, Pakistan, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, UK, United Arab Emirates, Philippines, and Russia