Event 88762

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
September 08, 2021
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Becton Dickinson Infusion Therapy Systems Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

BD Nexiva, 20 GA 1.00 IN (1.1 X 25 mm), 3660 ml/hr. 61ml/min; Closed IV Catheter System- Dual port -61 mL/min (3660 mL/hr)-0.5mL, REF 383536; BD Q-Style Luer Access Split Septum BD Vialon Material; BD Instaflash Needle Technology; Rx Only, Sterile EO. BD Nexiva closed IV catheter systems are over-the-needle, intravascular catheters. These devices have a radiopaque BD Vialon" catheter, needle, needle shield, septum, stabilization platform, integrated extension tubing, clamp, dual port adapter, vent plug and pre-attached BD Q-Syte Luer Access Split Septum. The needle and catheter are protected by a needle cover. A BD Q-Syte device or an end cap with protective cover is provided in the unit package. The closed system is designed to keep blood contained within the device throughout the insertion process. The septum is designed to wipe visible blood from the needle surface as the needle is withdrawn from the catheter, further reducing the risk of blood exposure. The needle tip is passively protected when the needle is removed, reducing the risk of accidental needlestick injury. These devices have BD Instaflash" Needle Technology, allowing for immediate visualization of blood along the catheter. Continuous blood return is seen in the extension tubing. The vent plug prevents blood leakage from the extension tubing during insertion. Both the stabilization platform and dual port adapter are color coded to indicate catheter gauge size (24GA (0.7 mm)=Yellow, 22GA (0.9 mm)=Blue, 20GA (1.1 mm)=Pink, 18GA (1.3 mm)=Green). BD Nexiva closed IV catheter systems are intended to be inserted into a patient s vascular system for short term use to sample blood, monitor blood pressure, or administer fluids. These catheters may be used for any patient population with consideration given to adequacy of vascular anatomy, procedure being performed, fluids being infused, and duration of therapy. The 22-18GA (0.9-1.3 mm) devices are suitable for use with power injectors set to a maximum pressure of 300 psi (2068 kPa) when access ports not suitable for use with power injectors are removed.

Z-0306-2022
Recall number
Z-0306-2022
Initiated
September 08, 2021
Classification
Class II
Status
Ongoing
Quantity
155,840 devices, 80 units in each box

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
There is a breach in the product packaging that renders the product non-sterile.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Packaging process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

There is a breach in the product packaging that renders the product non-sterile.

Code information

Catalog number/Model number: 383536; Batch numbers/UDI numbers: Lot: 1193055/(17)240630(10)1193055(30)80/(01)50382903835367, 2024-06-30; Lot: 1166273: (17)240630(10)1166273(30)80/(01)50382903835367, 2024-06-30 Lot: 1166273: (17)240630(10)1166273(30)80/(01)50382903835367, 2024-06-30 Lot: 1188953: (17)240630(10)1188953(30)80/(01)50382903835367, 2024-06-30

Distribution pattern

U.S. Nationwide Distribution: AR, AZ, MA, GA, NC, OH, WA, NY, NJ, HI, CA, MD, RI, MT, IN, PA, IL OK, OR, AK, CT, TX, ME, MO, FL, VA, LA, CO, UT O.U.S.: None