Recalling firm wording
FujiFilm Healthcare Americas Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Arietta 65 Ultrasound Software Versions: V1.0 through V4.0.0

Z-0192-2022
Recall number
Z-0192-2022
Initiated
October 14, 2021
Classification
Class II
Status
Ongoing
Quantity
48 US

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
When the ultrasound system is used in combination with a Fujifilm (formally Hitachi) transducer, there is an error in the calculation formula of Effective Regurgitant Orifice Area (EROA) when you perform MR Flow combined with Left Ventricular Outflow Tract Flow (LVOT Flow) after performing an echocardiogram. The incorrect value is displayed. In addition, the error appears in the instruction manual

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

When the ultrasound system is used in combination with a Fujifilm (formally Hitachi) transducer, there is an error in the calculation formula of Effective Regurgitant Orifice Area (EROA) when you perform MR Flow combined with Left Ventricular Outflow Tract Flow (LVOT Flow) after performing an echocardiogram. The incorrect value is displayed. In addition, the error appears in the instruction manual

Code information

Serial Numbers/UDI: G3024453 (01)04562122959278(11)191029(21)G3024453 G3024456 (01)04562122959278(11)191029(21)G3024456 G3042466 (01)04562122959278(11)191115(21)G3042466 G3042471 (01)04562122959278(11)191115(21)G3042471 G3042477 (01)04562122959278(11)191115(21)G3042477 G3042478 (01)04562122959278(11)191115(21)G3042478 G3042484 (01)04562122959278(11)191115(21)G3042484 G3042486 (01)04562122959278(11)191115(21)G3042486 G3042487 (01)04562122959278(11)191115(21)G3042487 G3042496 (01)04562122959278(11)191115(21)G3042496 G3042499 (01)04562122959278(11)191115(21)G3042499 G3042501 (01)04562122959278(11)191115(21)G3042501 G3042576 (01)04562122959278(11)191220(21)G3042576 G3042578 (01)04562122959278(11)191220(21)G3042578 G3042583 (01)04562122959278(11)191220(21)G3042583 G3042584 (01)04562122959278(11)191205(21)G3042584 G3042585 (01)04562122959278(11)191205(21)G3042585 G3042587 (01)04562122959278(11)191205(21)G3042587 G3042588 (01)04562122959278(11)191220(21)G3042588 G3082956 (01)04562122959278(11)200929(21)G3082956 G3085640 (01)04562122959278(11)200929(21)G3085640 G3085644 (01)04562122959278(11)200929(21)G3085644 G3095193 (01)04562122959278(11)201211(21)G3095193 G3095194 (01)04562122959278(11)201211(21)G3095194 G3095195 (01)04562122959278(11)201211(21)G3095195 G3095197 (01)04562122959278(11)201211(21)G3095197 G3095202 (01)04562122959278(11)201211(21)G3095202 G3095445 (01)04562122959278(11)201228(21)G3095445 G3111596 (01)04562122959278(11)210401(21)G3111596 G3111597 (01)04562122959278(11)210401(21)G3111597 G3111598 (01)04562122959278(11)210401(21)G3111598 G3111605 (01)04562122959278(11)210402(21)G3111605 G3111606 (01)04562122959278(11)210402(21)G3111606 G3114264 (01)04562122959278(11)210428(21)G3114264 G3114267 (01)04562122959278(11)210428(21)G3114267 G3114366 (01)04562122959278(11)210514(21)G3114366 G3114368 (01)04562122959278(11)210514(21)G3114368 G3114372 (01)04562122959278(11)210517(21)G3114372 G3114373 (01)04562122959278(11)210517(21)G3114373 G3114376 (01)04562122959278(11)210517(21)G3114376 G3114379 (01)04562122959278(11)210517(21)G3114379 G3114399 (01)04562122959278(11)210518(21)G3114399 G3114897 (01)04562122959278(11)210519(21)G3114897 G3114902 (01)04562122959278(11)210519(21)G3114902 G3118967 (01)04562122959278(11)210528(21)G3118967 G3118968 (01)04562122959278(11)210528(21)G3118968

Distribution pattern