device · product 1 of 1
Karl Storz Udel Sterilization Tray Instructions for Use, Tray Part Number/Description: 39301ACTS/CAMERA TRAY (STERRAD, INLINE CAMERA HEADS); 39301AS/RIGID TELESCOPE TRAY (STERRAD, 2 RIGID SCOPES 4MM X 18CM); 39301BCTS/CAMERA TRAY (INLINE/C-MOUNT CAMERA HEADS); 39301BS/RIGID TELESCOPE TRAY (STERRAD, 2 RIGID SCOPES 5MM X 35CM); 39301C1S/RIGID TELESCOPE TRAY (STERRAD, 1 RIGID SCOPE & LT CABLE); 39301CS/RIGID TELESCOPE TRAY (STERRAD, 2 RIGID SCOPE 10MM X 31CM); 39301DS/RIGID TELESCOPE TRAY (STERRAD, 2 RIGID SCOPES); 39301HCTS/CAMERA TRAY (STERRAD, STANDARD CAMERA HEADS); 39301PHTS/CAMERA TRAY (STERRAD, PENDULUM/INLINE CAM HEADS); 39305C1S/BASKET-STYLE RIGID TELESCOPE TRAY (UDEL, 1 RIGID SCOPE 4MM X 30CM); 39305C2S/BASKET-STYLE RIGID TELESCOPE TRAY (STERRAD, 2 RIGID SCOPES); 39305L1S/BASKET-STYLE RIGID TELESCOPE TRAY (STERRAD, 1 RIGID SCOPE); 39305L2S/BASKET-STYLE RIGID TELESCOPE TRAY (STERRAD, 2 RIGID SCOPES); 39311AS/RIGID TELESCOPE TRAY (STERRAD, 4 RIGID SCOPES & LT CABLE); 39311BS/RIGID TELESCOPE TRAY (STERRAD, TWO LAP SCOPES/CABLE); 39314FS/RIGID TELESCOPE TRAY (STERRAD, FOUR LAP SCOPES/CABLE); 39401AS/FLEXIBLE ENDOSCOPE TRAY (UP TO 39CM); 39402AS/FLEXIBLE ENDOSCOPE TRAY (FIBERSCOPES); 39403AS/FLEXIBLE ENDOSCOPE TRAY (CCD VIDESCOPES); 39406AS/FLEXIBLE ENDOSCOPE TRAY (CMOS VIDEOSCOPES)
- Recall number
- Z-0442-2022
- Initiated
- October 15, 2021
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Karl Storz Endoscopy
- Quantity
- 71892
App-derived interpretation
To ensure sterilization effectiveness, sterilization tray instructions for use were updated and now state 1) udel trays may be used with flexible endoscopes with lumen diameters greater-than-or-equal to 1.2mm and maximum length of 845mm and 2) ethylene oxide is excluded as a sterilization modality.
Official device-enrichment evidence · Sourced
Device Design
Inspect official wording and provenance
Reason for recall
To ensure sterilization effectiveness, sterilization tray instructions for use were updated and now state 1) udel trays may be used with flexible endoscopes with lumen diameters greater-than-or-equal to 1.2mm and maximum length of 845mm and 2) ethylene oxide is excluded as a sterilization modality.
Code information
All lots with Instructions for Use Prior to PI-000078-10.0 (06/21)
Distribution pattern
US: AK, AL, AP, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY OUS: Canada, Venezuela, Barbados, GUAM, Puerto Rico, Mexico, Northern Mariana Islands, Trinidad & Tobago, Germany, Cayman Islands, Philippines, Japan, Dominican Republic, Panama, Bahamas, Ecuador, Suriname, American Virgin Islands, British Virgin Islands, Belize, Curacao, The Bahamas, Brazil, South Korea, Bermuda