Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 88999

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
November 01, 2021
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Randox Laboratories Ltd.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

RANDOX CHOL-cholesterol (total) test system intended to measure cholesterol in plasma and serum REF: CH200

Z-0328-2022
Recall number
Z-0328-2022
Initiated
November 01, 2021
Classification
Class II
Status
Terminated
Recalling firm
Randox Laboratories Ltd.
Quantity
15 kits

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
When calibrating using the standard provided in the kit, the change in absorbance is lower than expected generating internal quality control results high outside range

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

When calibrating using the standard provided in the kit, the change in absorbance is lower than expected generating internal quality control results high outside range

Code information

Lot Numbers/Exp Date: 559344 28th June 2022; 572730 28th July 2023. GTIN: 05055273201130

Distribution pattern

US Nationwide distribution in the states of GA, OR, PA ,WA, West Virginia.