device · product 1 of 2
Baxter Spectrum IQ Infusion Pumps, Product code 3570009.
There is the potential for reduced or non-delivery of medication, in some cases without alerting the user via pump alarm. This may occur as a result of incorrect administration set setup and/or incomplete resolution of upstream occlusion alarms when using Spectrum V8 and Spectrum IQ infusion pumps.
These labels are deterministic app interpretations, not FDA categories.
There is the potential for reduced or non-delivery of medication, in some cases without alerting the user via pump alarm. This may occur as a result of incorrect administration set setup and/or incomplete resolution of upstream occlusion alarms when using Spectrum V8 and Spectrum IQ infusion pumps.
Code information
All serial numbers, GTIN 00085412610900.
Distribution pattern
Worldwide Distribution. US nationwide Puerto Rico and St. Thomas. There was government/military distribution. Countries of Canada, Bahamas, Barbados, Bermuda, Guyana, Jamaica, and Trinidad and Tobago.
device · product 2 of 2
Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 8), Product code 35700BAX2.
There is the potential for reduced or non-delivery of medication, in some cases without alerting the user via pump alarm. This may occur as a result of incorrect administration set setup and/or incomplete resolution of upstream occlusion alarms when using Spectrum V8 and Spectrum IQ infusion pumps.
These labels are deterministic app interpretations, not FDA categories.
There is the potential for reduced or non-delivery of medication, in some cases without alerting the user via pump alarm. This may occur as a result of incorrect administration set setup and/or incomplete resolution of upstream occlusion alarms when using Spectrum V8 and Spectrum IQ infusion pumps.
Code information
All serial numbers, GTIN 00085412498683.
Distribution pattern
Worldwide Distribution. US nationwide Puerto Rico and St. Thomas. There was government/military distribution. Countries of Canada, Bahamas, Barbados, Bermuda, Guyana, Jamaica, and Trinidad and Tobago.