Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 89035

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
November 10, 2021
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Beckman Coulter Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

SYNCHRON Systems Cholesterol (CHOL) Reagent, REF: 467825

Z-0399-2022
Recall number
Z-0399-2022
Initiated
November 10, 2021
Classification
Class II
Status
Ongoing
Recalling firm
Beckman Coulter Inc.
Quantity
1433

App-derived interpretation

Foreign material or chemical contamination reason.foreign_material_chemical_contamination · v1.0.0
particulates

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Cholesterol Reagent may fail with Out of Calibration Range Low (OCR Low) and/or Back to Back (B/B) errors. Low quality control recovery failures may be observed after passing calibration. Brown particulates in compartment C of the reagent cartridge may be observed. Continued use may lead to, 1) failed QC and/or calibrations, 2) Erroneous low Cholesterol patient results.

Code information

Lot: M106450, UDI: 15099590575113

Distribution pattern

US Nationwide distribution in the states of Louisiana, Ohio, North Carolina, California, South Carolina, Oklahoma, Pennsylvania, New York, Florida, New Jersey, Wisconsin, Alabama, Texas, Georgia, Maine, North Dakota, Virginia, Kansas, Washington, New Mexico, Iowa ,Nebraska, Idaho, Michigan, Maryland, Arizona, Hawaii, Missouri, Oregon, Mississippi, Rhode Island, Alaska, Tennessee, Illinois, Delaware, Kentucky, Arkansas, West Virginia, Massachusetts. OUS: India, South Africa, Saudi Arabia, Canada, Lebanon, Singapore, Philippines, Thailand