device · product 1 of 1
SYNCHRON Systems Cholesterol (CHOL) Reagent, REF: 467825
- Recall number
- Z-0399-2022
- Initiated
- November 10, 2021
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Beckman Coulter Inc.
- Quantity
- 1433
App-derived interpretation
particulates
Official device-enrichment evidence · Sourced
Nonconforming Material/Component
Inspect official wording and provenance
Reason for recall
Cholesterol Reagent may fail with Out of Calibration Range Low (OCR Low) and/or Back to Back (B/B) errors. Low quality control recovery failures may be observed after passing calibration. Brown particulates in compartment C of the reagent cartridge may be observed. Continued use may lead to, 1) failed QC and/or calibrations, 2) Erroneous low Cholesterol patient results.
Code information
Lot: M106450, UDI: 15099590575113
Distribution pattern
US Nationwide distribution in the states of Louisiana, Ohio, North Carolina, California, South Carolina, Oklahoma, Pennsylvania, New York, Florida, New Jersey, Wisconsin, Alabama, Texas, Georgia, Maine, North Dakota, Virginia, Kansas, Washington, New Mexico, Iowa ,Nebraska, Idaho, Michigan, Maryland, Arizona, Hawaii, Missouri, Oregon, Mississippi, Rhode Island, Alaska, Tennessee, Illinois, Delaware, Kentucky, Arkansas, West Virginia, Massachusetts. OUS: India, South Africa, Saudi Arabia, Canada, Lebanon, Singapore, Philippines, Thailand