Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 89144

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
December 03, 2015
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Smiths Medical ASD Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Medfusion Model 3500 and 4000 Syringe Infusion Pumps and Medfusion 3010 and 3010A Syringe Infusion Pumps serviced during the same time period in which a main board was replaced using replacement main board part numbers G6000361, G6000435, G6001260, G6001560, G6001562 and G6002728.

Z-0653-2022
Recall number
Z-0653-2022
Initiated
December 03, 2015
Classification
Class II
Status
Ongoing
Recalling firm
Smiths Medical ASD Inc.
Quantity
38,797 (31,065 pumps and 7,732 boards)

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
There have been intermittent issues with Supercap POST (power on self-test) and Backup Audible Alarm error codes outside of the pump warranty period. These alarms occur at start up when the pump goes through the self-test. This recall occurred in 2015.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Component change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

There have been intermittent issues with Supercap POST (power on self-test) and Backup Audible Alarm error codes outside of the pump warranty period. These alarms occur at start up when the pump goes through the self-test. This recall occurred in 2015.

Code information

Medfusion 3500 and 4000 Syringe Infusion Pumps manufactured between April 2012 and June 2015; and Medfusion 3010 and 3010A (and pump models 3500 and 40000 serviced during April 20121 and June 2015 in which a main board was replaced.

Distribution pattern

Distribution US nationwide, including PR, VI, GU, and MP. There was government/military/foreign distribution.