Alinity m System, Part No. 08N53-002
- Recall number
- Z-0462-2022
- Initiated
- December 06, 2021
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Abbott Molecular, Inc.
- Quantity
- 742 devices
App-derived interpretation
software defect
Official device-enrichment evidence · Sourced
Software design
Inspect official wording and provenance
Reason for recall
There is a software defect that can cause the Amplification Detection clamp CLEAN position to be set too high.
Code information
Alinity m System, Model 08N53-002, all serial numbers
Distribution pattern
Worldwide distribution - US distribution and the countries of Australia, Austria, Belgium, Brazil, Cambodia, Canada, Chile, Colombia, Czech Republic, Egypt, El Salvador, Estonia, Finland, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Malawi, Malaysia, Mexico, Montenegro, Morocco, Mozambique, Netherlands, New Zealand, Norway, Pakistan, Poland, Portugal, Romania, Russia, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Uganda, United Kingdom, and Vietnam