Philips Azurion Interventional Fluoroscopic X-ray System with Azurion R2.2 with LUC configuration start-up issue; Azurion R2.2 emergency stop recovery issue Model Numbers: 722221, 722228, 722225, 722068, 722226, 722078, 722224
- Recall number
- Z-0471-2022
- Initiated
- November 12, 2021
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Philips North America Llc
- Quantity
- 10 units (1 US and 9 OUS)
App-derived interpretation
(1)Start-up problem: Intermittently at start-up of the system, the communication of the control unit that manages the movement of the stand is not established; (2) Emergency stop recovery problem-warm restart is not executed correctly and movement functionality is not available
Official device-enrichment evidence · Sourced
Software design
Inspect official wording and provenance
Reason for recall
(1)Start-up problem: Intermittently at start-up of the system, the communication of the control unit that manages the movement of the stand is not established; (2) Emergency stop recovery problem-warm restart is not executed correctly and movement functionality is not available
Code information
US: Model 722068 Azurion 7 B20 S/N: 109 UDI: (01)00884838085367(21)109. OUS: Model: S/N UDI: 722226 Azurion 7 B20 176 (01)00884838099272(21)176; 722228 Azurion 5 M20 3 (01)00884838099234(21)3; 722221 Azurion 3 M12 8 (01)00884838099203(21)8; 722225 Azurion 7 B12 16 (01)00884838099265(21)16; 722225 Azurion 7 B12 73 (01)00884838099265(21)73; 722226 Azurion 7 B20 42 (01)00884838099272(21)42; 722226 Azurion 7 B20 157 (01)00884838099272(21)157; 722224 Azurion 7 M20 293 (01)00884838099258(21)293; 722078 Azurion 7 M12 220 (01)00884838085251(21)220
Distribution pattern
Worldwide distribution - US Nationwide distribution in the state of WA and the countries of France, Germany, Japan, Poland, United Kingdom.