Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 89235

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
December 15, 2021
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
MEDLINE INDUSTRIES, LP - Northfield

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

CURAD REUSABLE GEL HOT/COLD PACK

Z-0633-2022
Recall number
Z-0633-2022
Initiated
December 15, 2021
Classification
Class II
Status
Ongoing
Quantity
494,273 ea

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The product contains natural rubber latex but the label states that it does not.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The product contains natural rubber latex but the label states that it does not.

Code information

Item Numbers: 1) CUR959/CUR959H (Lots 69718070004, 69718090003, 69718100001, 69718100004, 69718110003, 69719010001, 69719030001, 69719030004, 69719040002, 69719070005, 69719080002, 69719080006, 69719090001, 69719090004, 69719100001, 69719100004, 69719110001, 69719110003, 69719110004, 69719120001, 69719120005, 69719120002, 69719120006, 69720010001, 69720020002, 69720040001, 69720050001, 69720060001, 69720060002, 69720080002, 69720090001, 69720090002, 69720100001, 69720100002, 69720110002, 69720120001, 69720120002, 69721010001, 69721020001, 69721030001, 69721030002, 69721050001, 69721050002, 69721050003, 69721060001, 69721060002, 69721070001, 69721080002, 69721090001) 2) TGT0114V1 (Lots 69718070004, 69718080002, 69718090002 , 69718100004, 69718110003, 69719010001, 69719030004, 69719040002, 69719050002, 69719050004, 69719060001, 69719060003, 69719070001, 69719070005, 69719080002, 69719080006, 69719090004, 69719100001, 69719100004, 69719100005, 69719110001, 69719110002, 69719110004, 69719110003, 69719120005, 69719120002, 69719120007, 69719120006, 69720010001, 69720070001, 69720070002, 69720080001, 69720080002, 69720100003)

Distribution pattern

US and Bahamas, Bermuda, Panama, Lebanon, Barbados, Honduras, Anguilla