Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 89338

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
January 04, 2022
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
BALT USA, LLC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

The Optima Coil System consists of an implantable embolization coil comprised of a platinum-tungsten alloy attached to a proximal stainless steel hypo-tube and distal body coil delivery pusher with a radiopaque distal positioning marker and proximal fluorosafe marker. The Optima Coil delivery pusher is 185cm in length. it is designed for use with the XCEL Detachment Controller. The XCEL Detachment Controller comes pre-loaded with batteries and is a sterile, handheld, single-patient-use device which should not to be reused, re-sterilized, opened or tampered with. The Optima Coil System and XCEL Detachment Controller(s) are sold separately. The Optima Coil achieves detachment by an internal heater element, which is powered by the XCEL Detachment Controller.

Z-0660-2022
Recall number
Z-0660-2022
Initiated
January 04, 2022
Classification
Class II
Status
Ongoing
Recalling firm
BALT USA, LLC
Quantity
51 devices

App-derived interpretation

Labeling or packaging reason.labeling_packaging · v1.0.0
Label/Pouch mix-up

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Due to Label/Pouch mix-up. Pouches incorrectly labeled. Incorrect shelf carton label or pouch label due to inadequate manufacturing line clearance. As a result of this labeling mix-up, there is the possibility of inadvertently selecting the incorrect coil size for the procedure potentially causing some harms like vessel damage, an inability to treat with said device and potential rupture of the aneurysm.

Code information

Model/REF: OPTI0204CSF10 Lot Number: F210500134 UDI Code: 00818053026010 Model/REF: OPTI0407CSF10 Lot Number: F210500135 UDI Code: 00818053026096

Distribution pattern

Worldwide distribution - U.S. Nationwide distribution in the states of FL and MD. The countries of Denmark, France, Germany, Lebanon, Netherlands, Sweden, Peru, Portugal, Slovenia, Spain, and United Arab Emirates.