Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 89509

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
November 01, 2021
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Human Care USA, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

Hanger-Bar 2P 45cm

Z-0851-2022
Recall number
Z-0851-2022
Initiated
November 01, 2021
Classification
Class II
Status
Ongoing
Recalling firm
Human Care USA, Inc.
Quantity
10 devices

App-derived interpretation

Manufacturing or process control reason.manufacturing_process_control · v1.0.0
manufacturing issue (failure

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The Hanger-Bar is being recalled due to the potential for the set screw loosening as a result of a manufacturing issue (failure to apply Loctite adhesive). As a result of the set screw loosening, there is the potential for a fall hazard.

Code information

Part Number 72754 UPC Code: 881608800397 EAN Code: 7331599999707 Lot Number: 33218501

Distribution pattern

U.S. Nationwide distribution in the states of CA, IL, LA, MA, NV, OH, OK, PA, TX, UT, VA, and WV.

device · product 2 of 2

Hanger-Bar 2P 45cm

Z-0852-2022
Recall number
Z-0852-2022
Initiated
November 01, 2021
Classification
Class II
Status
Ongoing
Recalling firm
Human Care USA, Inc.
Quantity
15 devices

App-derived interpretation

Manufacturing or process control reason.manufacturing_process_control · v1.0.0
manufacturing issue (failure

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The Hanger-Bar is being recalled due to the potential for the set screw loosening as a result of a manufacturing issue (failure to apply Loctite adhesive). As a result of the set screw loosening, there is the potential for a fall hazard.

Code information

Part Number 72755 UPC Code: 881608800403 EAN Code: 7331599999714 Lot Number: 33164504

Distribution pattern

U.S. Nationwide distribution in the states of CA, IL, LA, MA, NV, OH, OK, PA, TX, UT, VA, and WV.