Event 89544

8 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
January 07, 2022
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Biomet, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

8 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 8

Arcos Modular Revision Hip System, Standard Cone Prox Body, Porous Plasma,Size A, 60 MM Item Number: 11-301301

Z-0670-2022
Recall number
Z-0670-2022
Initiated
January 07, 2022
Classification
Class II
Status
Terminated
Recalling firm
Biomet, Inc.
Quantity
19 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor.

Code information

Lot Number: 853650, 853810 UDI: (01)00880304468030(17) 290122(10)853650; (01)00880304468030(17) 290211(10)853810

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Canada, Ecuador, Netherlands.

device · product 2 of 8

Comprehensive Shoulder System Primary Shoulder Stem, Micro Length, Porous Plasma, 11 MM, 55 MM Long Item Number: 113611

Z-0671-2022
Recall number
Z-0671-2022
Initiated
January 07, 2022
Classification
Class II
Status
Terminated
Recalling firm
Biomet, Inc.
Quantity
10 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor.

Code information

Lot Number: 821170 UDI: (01)00880304554122(17) 280825(10)821170

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Canada, Ecuador, Netherlands.

device · product 3 of 8

Comprehensive Shoulder System Primary Shoulder Stem, Micro Length, Porous Plasma, 14 MM, 55 MM Long Item Number: 113614

Z-0672-2022
Recall number
Z-0672-2022
Initiated
January 07, 2022
Classification
Class II
Status
Terminated
Recalling firm
Biomet, Inc.
Quantity
12 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor.

Code information

Lot Number: 439470 UDI: (01)00880304554153(17) 290803(10)439470

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Canada, Ecuador, Netherlands.

device · product 4 of 8

Comprehensive Shoulder System Primary Shoulder Stem, Micro Length, Porous Plasma, 18 MM, 55 MM Long Item Number: 113618

Z-0673-2022
Recall number
Z-0673-2022
Initiated
January 07, 2022
Classification
Class II
Status
Terminated
Recalling firm
Biomet, Inc.
Quantity
12 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor.

Code information

Lot Number: 732370 UDI: (01)00880304554191(17) 290802(10)732370

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Canada, Ecuador, Netherlands.

device · product 5 of 8

Comprehensive Shoulder System Mini Humeral Stem, Porous Plasma, 11 MM, 83 MM Long Item Number: 113631

Z-0674-2022
Recall number
Z-0674-2022
Initiated
January 07, 2022
Classification
Class II
Status
Terminated
Recalling firm
Biomet, Inc.
Quantity
11 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor.

Code information

Lot Number: 821860 UDI: (01)00880304462687(17) 290522(10)821860

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Canada, Ecuador, Netherlands.

device · product 6 of 8

Comprehensive Shoulder System Mini Humeral Stem, Porous Plasma, 12 MM, 83 MM Long Item Number: 113632

Z-0675-2022
Recall number
Z-0675-2022
Initiated
January 07, 2022
Classification
Class II
Status
Terminated
Recalling firm
Biomet, Inc.
Quantity
12 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor.

Code information

Lot Number: 327160 UDI: (01)00880304449749(17)290819(10)327160

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Canada, Ecuador, Netherlands.

device · product 7 of 8

Comprehensive Shoulder System Mini Humeral Stem, Porous Plasma, 13 MM, 83 MM Long Item Number: 113633

Z-0676-2022
Recall number
Z-0676-2022
Initiated
January 07, 2022
Classification
Class II
Status
Terminated
Recalling firm
Biomet, Inc.
Quantity
9 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor.

Code information

Lot Number: 994800 UDI:(01)00880304462632(17)290816(10)994800

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Canada, Ecuador, Netherlands.

device · product 8 of 8

Comprehensive Shoulder System Mini Humeral Stem, Porous Plasma, 15 MM, 83 MM Long Item Number: 113635

Z-0677-2022
Recall number
Z-0677-2022
Initiated
January 07, 2022
Classification
Class II
Status
Terminated
Recalling firm
Biomet, Inc.
Quantity
11 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor.

Code information

Lot Number: 396250 UDI: (01)00880304462663(17)281121(10)396250

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Canada, Ecuador, Netherlands.