Alinity s System software version2.8.0, REF LN 06P16-01, IVD, OEI: (01)00380740138479 The Alinity s System is intended for In Vitro diagnostic use only. The Alinity s System is a highthroughput, fully automated immunoassay analyzer designed to determine the presence of specific antigens and antibodies by using chemiluminescent immunoassay technology.
- Recall number
- Z-0749-2022
- Initiated
- February 03, 2022
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Abbott Laboratories, Inc
- Quantity
- 306 units
App-derived interpretation
Software error
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Inspect official wording and provenance
Reason for recall
Software error associated with the immunoassay analyzer wash cycle which is using 1 mL of wash buffer instead of the intended 3 mL of wash buffer to wash the exterior of the probe.
Code information
Serial numbers AS1001 through AS1387 UDIs: (01)00380740138479(21)AS1002 (01)00380740138479(21)AS1005 (01)00380740138479(21)AS1009 (01)00380740138479(21)AS1012 (01)00380740138479(21)AS1013 (01)00380740138479(21)AS1016 (01)00380740138479(21)AS1019 (01)00380740138479(21)AS1023 (01)00380740138479(21)AS1028 (01)00380740138479(21)AS1031 (01)00380740138479(21)AS1032 (01)00380740138479(21)AS1044 (01)00380740138479(21)AS1058 (01)00380740138479(21)AS1064 (01)00380740138479(21)AS1097 (01)00380740138479(21)AS1126 (01)00380740138479(21)AS1148 (01)00380740138479(21)AS1153 (01)00380740138479(21)AS1155 (01)00380740138479(21)AS1160 (01)00380740138479(21)AS1161 (01)00380740138479(21)AS1193 (01)00380740138479(21)AS1212 (01)00380740138479(21)AS1213 (01)00380740138479(21)AS1215 (01)00380740138479(21)AS1217 (01)00380740138479(21)AS1218 (01)00380740138479(21)AS1219 (01)00380740138479(21)AS1220 (01)00380740138479(21)AS1221 (01)00380740138479(21)AS1222 (01)00380740138479(21)AS1224 (01)00380740138479(21)AS1225 (01)00380740138479(21)AS1226 (01)00380740138479(21)AS1227 (01)00380740138479(21)AS1230 (01)00380740138479(21)AS1231 (01)00380740138479(21)AS1233 (01)00380740138479(21)AS1234 (01)00380740138479(21)AS1235 (01)00380740138479(21)AS1236 (01)00380740138479(21)AS1237 (01)00380740138479(21)AS1238 (01)00380740138479(21)AS1240 (01)00380740138479(21)AS1241 (01)00380740138479(21)AS1242 (01)00380740138479(21)AS1243 (01)00380740138479(21)AS1246 (01)00380740138479(21)AS1247 (01)00380740138479(21)AS1248 (01)00380740138479(21)AS1249 (01)00380740138479(21)AS1250 (01)00380740138479(21)AS1251 (01)00380740138479(21)AS1252 (01)00380740138479(21)AS1253 (01)00380740138479(21)AS1254 (01)00380740138479(21)AS1255 (01)00380740138479(21)AS1256 (01)00380740138479(21)AS1257 (01)00380740138479(21)AS1259 (01)00380740138479(21)AS1261 (01)00380740138479(21)AS1262 (01)00380740138479(21)AS1263 (01)00380740138479(21)AS1264 (01)00380740138479(21)AS1266 (01)00380740138479(21)AS1267 (01)00380740138479(21)AS1268 (01)00380740138479(21)AS1269 (01)00380740138479(21)AS1270 (01)00380740138479(21)AS1271 (01)00380740138479(21)AS1272 (01)00380740138479(21)AS1273 (01)00380740138479(21)AS1274 (01)00380740138479(21)AS1275 (01)00380740138479(21)AS1276 (01)00380740138479(21)AS1277 (01)00380740138479(21)AS1278 (01)00380740138479(21)AS1279 (01)00380740138479(21)AS1280 (01)00380740138479(21)AS1281 (01)00380740138479(21)AS1282 (01)00380740138479(21)AS1283 (01)00380740138479(21)AS1284 (01)00380740138479(21)AS1285 (01)00380740138479(21)AS1286 (01)00380740138479(21)AS1287 (01)00380740138479(21)AS1288 (01)00380740138479(21)AS1289 (01)00380740138479(21)AS1290 (01)00380740138479(21)AS1291 (01)00380740138479(21)AS1293 (01)00380740138479(21)AS1294 (01)00380740138479(21)AS1295 (01)00380740138479(21)AS1296 (01)00380740138479(21)AS1299 (01)00380740138479(21)AS1300 (01)00380740138479(21)AS1301 (01)00380740138479(21)AS1302 (01)00380740138479(21)AS1303 (01)00380740138479(21)AS1304 (01)00380740138479(21)AS1305 (01)00380740138479(21)AS1306 (01)00380740138479(21)AS1307 (01)00380740138479(21)AS1309 (01)00380740138479(21)AS1310 (01)00380740138479(21)AS1311 (01)00380740138479(21)AS1312 (01)00380740138479(21)AS1313 (01)00380740138479(21)AS1314 (01)00380740138479(21)AS1315 (01)00380740138479(21)AS1316 (01)00380740138479(21)AS1317 (01)00380740138479(21)AS1318 (01)00380740138479(21)AS1319 (01)00380740138479(21)AS1320 (01)00380740138479(21)AS1321 (01)00380740138479(21)AS1322 (01)00380740138479(21)AS1323 (01)00380740138479(21)AS1325 (01)00380740138479(21)AS1326 (01)00380740138479(21)AS1327 (01)00380740138479(21)AS1330 (01)00380740138479(21)AS1332 (01)00380740138479(21)AS1333 (01)00380740138479(21)AS1334 (01)00380740138479(21)AS1336 (01)00380740138479(21)AS1341 (01)00380740138479(21)AS1344 (01)00380740138479(21)AS1346 (01)00380740138479(21)AS1347 (01)00380740138479(21)AS1348 (01)00380740138479(21)AS1349 (01)00380740138479(21)AS1350 (01)00380740138479(21)AS1351 (01)00380740138479(21)AS1352 (01)00380740138479(21)AS1353 (01)00380740138479(21)AS1354 (01)00380740138479(21)AS1355 (01)00380740138479(21)AS1358 (01)00380740138479(21)AS1359 (01)00380740138479(21)AS1360 (01)00380740138479(21)AS1368
Distribution pattern
Worldwide distribution: US (nationwide) including states of: AL, CA, GA, IL, IN, MN, MO, NC, OK, PA, RI, SC, TN, TX, WA, WI, MD and countries (OUS) including of: Australia, Austria, Bahrain, Belgium, Brazil, china, Colombia, Croatia, Cyprus, Czech Republic, Egypt, Germany, Greece, Hong Kong, Hungary, Iran, Ireland, Israel, Italy, Jordan Kuwait and Mexico.