Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 89647

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
March 01, 2022
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Edwards Lifesciences, LLC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

FORE-SIGHT ELITE Absolute Tissue Oximeter, REF: 01-06-3000, Contents: Monitor 1 EA., Accessory Kit 1 EA.

Z-0834-2022
Recall number
Z-0834-2022
Initiated
March 01, 2022
Classification
Class II
Status
Ongoing
Recalling firm
Edwards Lifesciences, LLC
Quantity
402 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The StO2 values may be inaccurately low when using either the FORE-SIGHT ELITE Tissue Oximeter Module or the FORE-SIGHT ELITE Absolute Tissue Oximeter Monitor with the Fore-Sight Elite large sensor in certain somatic locations (arms and legs). While the StO2 absolute values are impacted, the directional trend remains accurate, but may have a larger magnitude change. Low StO2 values may lead to unintended or inappropriate treatment

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software in the Use Environment

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The StO2 values may be inaccurately low when using either the FORE-SIGHT ELITE Tissue Oximeter Module or the FORE-SIGHT ELITE Absolute Tissue Oximeter Monitor with the Fore-Sight Elite large sensor in certain somatic locations (arms and legs). While the StO2 absolute values are impacted, the directional trend remains accurate, but may have a larger magnitude change. Low StO2 values may lead to unintended or inappropriate treatment

Code information

FORE-SIGHT ELITE Tissue Oximeter, Model Number 01-06-3000, UDI: 10609538630009; All Serial numbers.

Distribution pattern

Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Brunei, Bulgaria, Canada, CANARY ISLANDS, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, Estonia, Finland, France, Germany, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Iran, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Malaysia, Mexico, Netherlands, New Zealand, Northern Ireland, Norway, Oman, Panama, Peru, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates (UAE), United Kingdom, Vietnam.

device · product 2 of 2

FORE-SIGHT ELITE Tissue Oximeter module, REF: HEMFSM10

Z-0835-2022
Recall number
Z-0835-2022
Initiated
March 01, 2022
Classification
Class II
Status
Ongoing
Recalling firm
Edwards Lifesciences, LLC
Quantity
8389 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The StO2 values may be inaccurately low when using either the FORE-SIGHT ELITE Tissue Oximeter Module or the FORE-SIGHT ELITE Absolute Tissue Oximeter Monitor with the Fore-Sight Elite large sensor in certain somatic locations (arms and legs). While the StO2 absolute values are impacted, the directional trend remains accurate, but may have a larger magnitude change. Low StO2 values may lead to unintended or inappropriate treatment

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software in the Use Environment

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The StO2 values may be inaccurately low when using either the FORE-SIGHT ELITE Tissue Oximeter Module or the FORE-SIGHT ELITE Absolute Tissue Oximeter Monitor with the Fore-Sight Elite large sensor in certain somatic locations (arms and legs). While the StO2 absolute values are impacted, the directional trend remains accurate, but may have a larger magnitude change. Low StO2 values may lead to unintended or inappropriate treatment

Code information

FORE-SIGHT ELITE Tissue Oximeter module, REF: HEMFSM10, UDI: 00690103208573; All serial numbers.

Distribution pattern

Worldwide distribution - US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Brunei, Bulgaria, Canada, CANARY ISLANDS, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, Estonia, Finland, France, Germany, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Iran, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Malaysia, Mexico, Netherlands, New Zealand, Northern Ireland, Norway, Oman, Panama, Peru, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates (UAE), United Kingdom, Vietnam.