device · product 1 of 1
ExacTrac Dynamic software model numbers: 20910-01, ETD Positioning and Moni. 1.0.0 INST; 20910-01A, ETD Positioning and Moni. 1.0.1 INST; 20910-01B, ETD Positioning and Moni. 1.0.2 INST; and 20910-01C ETD Positioning and Moni. 1.0.3 INST.
- Recall number
- Z-0733-2022
- Initiated
- February 25, 2022
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Brainlab AG
- Quantity
- 205 systems
App-derived interpretation
The yaw angle may be incorrect for CBCT positioning workflows using setup beams with Varian LINACs.
Official device-enrichment evidence · Sourced
Software design
Inspect official wording and provenance
Reason for recall
The yaw angle may be incorrect for CBCT positioning workflows using setup beams with Varian LINACs.
Code information
Software versions 1.0.0, 1.0.1, 1.0.2, 1.0.3, GTIN 04056481142315.
Distribution pattern
Distribution was made to AZ, AR, CA, CO, FL, GA, IN, KY, MD, MS, MO, NJ, NY, NC, OK, OR, PA, TX, VA, and WA. There was no government or military distribution. Foreign distribution was made to Australia, Austria, Bahrain, Belgium, China, Denmark, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Jordan, New Zealand, Poland, Singapore, Slovakia, South Korea, Spain, Thailand, Turkey, United Kingdom, and United Arab Emirates.