device · product 1 of 1
Illumina NextSeq 550Dx, REF: 20005715, CE, IVD
- Recall number
- Z-0854-2022
- Initiated
- February 22, 2022
- Classification
- Class II
- Status
- Completed
- Recalling firm
- Illumina, Inc.
- Quantity
- 674 instruments
App-derived interpretation
There is a potential that the DNA sequence analyzer may short circuit shortly after disconnection and reconnection of circuit board or imaging module due to an insufficient reconnection of the ribbon cable. The short circuit results in the device being inoperable.
Official device-enrichment evidence · Sourced
Device Design
Inspect official wording and provenance
Reason for recall
There is a potential that the DNA sequence analyzer may short circuit shortly after disconnection and reconnection of circuit board or imaging module due to an insufficient reconnection of the ribbon cable. The short circuit results in the device being inoperable.
Code information
Lot #s NDX550110 - NDX550843 ; UDI: (01)00816270020125
Distribution pattern
US: PA, WA, CA, MI, FL, MD, AR, UT NC, Fl, IL, NE, NJ, TN, NY, IN, CO, GA, OK, TX, CT, SD, NM, VA, AL, LA, SC, WV, OR, IA, AZ, MN OUS: Australia Austria Belgium Chile China Denmark Finland France Germany Hungary Ireland Italy Japan Latvia Netherlands Norway Poland Portugal Romania Russian Federation Saudi Arabia Singapore Slovakia South Africa South Korea Spain Sweden Switzerland Turkey United Arab Emirates United Kingdom Vietnam