Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 89750

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
February 02, 2022
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Philips Healthcare

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Philips StentBoost Live R2.0 application, when used with the Philips Azurion with software release R2.x

Z-0737-2022
Recall number
Z-0737-2022
Initiated
February 02, 2022
Classification
Class II
Status
Terminated
Recalling firm
Philips Healthcare
Quantity
42 Systems

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Due to an incorrect configuration setting in the EPX database of StentBoost Live, acquisition does not stop automatically as it should. When the user selects the StentBoost Live protocol on the Philips Azurion system, the following on-screen message is displayed to the user in the Philips StentBoost Live application: "Press the cine pedal until the acquisition stops". However, due to the incorrect configuration of the EPX database, the acquisition does not stop after 40 images, but instead it continues as long as the pedal is pressed.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Radiation Control for Health and Safety Act

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Due to an incorrect configuration setting in the EPX database of StentBoost Live, acquisition does not stop automatically as it should. When the user selects the StentBoost Live protocol on the Philips Azurion system, the following on-screen message is displayed to the user in the Philips StentBoost Live application: "Press the cine pedal until the acquisition stops". However, due to the incorrect configuration of the EPX database, the acquisition does not stop after 40 images, but instead it continues as long as the pedal is pressed.

Code information

Azurion R2.1

Distribution pattern