Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 89843

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
March 14, 2022
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Neuralynx Inc

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

ATLAS Stim Headbox-NK For use only with the ATLAS Neurophysiology System. CAUTION--INVESTIGATIONAL DEVICE Limited by Federal (or United States) law to investigational use. Only for use in Non-Significant Risk studies under the supervision of an IRB.

Z-1091-2022
Recall number
Z-1091-2022
Initiated
March 14, 2022
Classification
Class II
Status
Ongoing
Recalling firm
Neuralynx Inc
Quantity
15

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Atlas Stim Headbox (ASHB) was delivered to US customers under a premise that the device had an Investigational Device Exemption (IDE) and was for research use only, but the manufacturer does not have an approved IDE by the FDA. The product has not undergone electrical safety testing and has not received regulatory clearance.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

No Marketing Application

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Atlas Stim Headbox (ASHB) was delivered to US customers under a premise that the device had an Investigational Device Exemption (IDE) and was for research use only, but the manufacturer does not have an approved IDE by the FDA. The product has not undergone electrical safety testing and has not received regulatory clearance.

Code information

ASHB Part No. 31-0601-0077 (SN0001, SN0002, SN0004, SN0005, SN0006, SN0007, SN0008, SN0009, SN0010, SN0012); ASHB Part No. 31-0601-0132(SN0003, SN0005, SN0006); ASHB Part No. 31-0601-0089(SN0004, SN0006);

Distribution pattern

U.S. Nationwide distribution in the states of AL, CA, IL, MN, NC, RI.