device · product 1 of 4
syngo.via RT Image Suite upgraded to VB60 running on syngo.via Model Number 10496180. Running on servers with syngo.via VB60 software after upgrade and CT scanners with syngo.CT VA40A after upgrade.
- Recall number
- Z-0994-2022
- Initiated
- March 07, 2022
- Classification
- Class II
- Status
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Quantity
- 1 (US)
App-derived interpretation
software version makes an automatic change in laser configuration settings for "markerless workflow" which may
Official device-enrichment evidence · Sourced
Software change control
Inspect official wording and provenance
Reason for recall
Upgraded software version makes an automatic change in laser configuration settings for "markerless workflow" which may lead to the wrong laser offset coordinates being displayed in the Patient Marking step.
Code information
Serial Number 133527
Distribution pattern
Worldwide distribution - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, IL, IN, KY, MD, MO, ND, NH, NY, OH, OR, PA, SD, TX, VA and the countries of Argentina, Austria, Bahrain, Belgium, Brazil, Czechia, Denmark, France, Germany, Hungary, India, Italy, Japan, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Switzerland, United Kingdom.