Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 89959

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
March 11, 2022
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
DePuy Orthopaedics, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

VELYS Robotic-Assisted Solution Base Product No.: 451570100

Z-1029-2022
Recall number
Z-1029-2022
Initiated
March 11, 2022
Classification
Class II
Status
Terminated
Recalling firm
DePuy Orthopaedics, Inc.
Quantity
US: 68 units; OUS: 10 units

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
software v1.5.1 has a system software issue

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Software design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

System software v1.5.1 has a system software issue related to the Daylight Savings Time (DST) change that can cause a system error, requiring the user to restart the system and potentially cause a delay in treatment

Code information

Serial Numbers (US): BAS1111 BAS1145 BAS1154 BAS1069 BAS1070 BAS1125 BAS1144 BAS1174 BAS1175 BAS1192 BAS1194 BAS1081 BAS1082 BAS1052 BAS1165 BAS1136 BAS1071 BAS1142 BAS1073 BAS1189 BAS1022 BAS1190 BAS1079 BAS1054 BAS1061 BAS1173 BAS1204 BAS1141 BAS1076 BAS1181 BAS1176 BAS1184 BAS1119 BAS1137 BAS1009 BAS1140 BAS1121 BAS1143 BAS1127 BAS1060 BAS1004 BAS1196 BAS1120 BAS1068 BAS1152 BAS1188 BAS1078 BAS1195 BAS1191 BAS1062 BAS1180 BAS1118 BAS1197 BAS1139 BAS1066 BAS1126 BAS1093 BAS1122 BAS1134 BAS1203 BAS1132 BAS1185 BAS1182 BAS1129 BAS1063 BAS1183 BAS1133 BAS1212 UDI: 10603295519515

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Israel, Australia & New Zealand.