Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 90122

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
June 17, 2022
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Luminex Corporation

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

Verigene CDF Nucleic Acid Test, Part No. 30-002-22

Z-1334-2022
Recall number
Z-1334-2022
Initiated
June 17, 2022
Classification
Class II
Status
Ongoing
Recalling firm
Luminex Corporation
Quantity
296 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
There is a potential for false negative results using VERIGENE CDF Stool PREP KIT and VERIGENE EP Stool PREP KIT due to hydrophobic characteristics of the swab.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

There is a potential for false negative results using VERIGENE CDF Stool PREP KIT and VERIGENE EP Stool PREP KIT due to hydrophobic characteristics of the swab.

Code information

UDI 00840487101735 Luminex Lot Number 042021022D, FLOQ Swab Lot Numbers: 2100856, 2032870; Luminex Lot Number 071321022F, FLOQ Swab Lot Numbers: 2100856, 2107946; Luminex Lot Number 100421022A, FLOQ Swab Lot Numbers: 2110850; Luminex Lot Number 110221022E, FLOQ Swab Lot Numbers: 2107946.

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Kuwait, Austria, France, Greece, Kuwait, Turkey.

device · product 2 of 2

VERIGENE Enteric Pathogens Nucleic Acid Test, Part No. 30-002-23

Z-1335-2022
Recall number
Z-1335-2022
Initiated
June 17, 2022
Classification
Class II
Status
Ongoing
Recalling firm
Luminex Corporation
Quantity
5601 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
There is a potential for false negative results using VERIGENE CDF Stool PREP KIT and VERIGENE EP Stool PREP KIT due to hydrophobic characteristics of the swab.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Nonconforming Material/Component

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

There is a potential for false negative results using VERIGENE CDF Stool PREP KIT and VERIGENE EP Stool PREP KIT due to hydrophobic characteristics of the swab.

Code information

UDI 00840487101568 Luminex Lot Number 030121023A, FLOQSwab Lot Number: 2100856; Luminex Lot Number 030221023D, FLOQSwab Lot Number: 2100856; Luminex Lot Number 030221023E, FLOQSwab Lot Number: 2100856; Luminex Lot Number 030221023F, FLOQSwab Lot Number: 2100856; Luminex Lot Number 030821023A, FLOQSwab Lot Number: 2100856; Luminex Lot Number 030921023D, FLOQSwab Lot Number: 2100856; Luminex Lot Number 031021023E, FLOQSwab Lot Number: 2100856; Luminex Lot Number 031621023D, FLOQSwab Lot Number: 2100856; Luminex Lot Number 031721023D, FLOQSwab Lot Number: 2100856; Luminex Lot Number 032221023A, FLOQSwab Lot Number: 2032870; Luminex Lot Number 032421023E, FLOQSwab Lot Number: 2032870; Luminex Lot Number 032521023E, FLOQSwab Lot Number: 2100856; Luminex Lot Number 033021023E, FLOQSwab Lot Number: 2100856; Luminex Lot Number 041421023F, FLOQSwab Lot Number: 2100856; Luminex Lot Number 050521023H, FLOQSwab Lot Number: 2100856; Luminex Lot Number 050621023C, FLOQSwab Lot Number: 2032870; Luminex Lot Number 051221023E, FLOQSwab Lot Number: 2100856; Luminex Lot Number 051221023F, FLOQSwab Lot Number: 2100856, 2032870; Luminex Lot Number 051721023B, FLOQSwab Lot Number: 2100856, 2032870; Luminex Lot Number 052721023D, FLOQSwab Lot Number: 2107946; Luminex Lot Number 060121023A, FLOQSwab Lot Number: 2107946; Luminex Lot Number 060921023C, FLOQSwab Lot Number: 2032870; Luminex Lot Number 062321023E, FLOQSwab Lot Number: 2107946; Luminex Lot Number 062921023D, FLOQSwab Lot Number: 2107946; Luminex Lot Number 063021023D, FLOQSwab Lot Number: 2110850; Luminex Lot Number 063021023E, FLOQSwab Lot Number: 2107946; Luminex Lot Number 070121023D, FLOQSwab Lot Number: 2107946; Luminex Lot Number 070621023A, FLOQSwab Lot Number: 2107946; Luminex Lot Number 070821023G, FLOQSwab Lot Number: 2107946; Luminex Lot Number 070821023H, FLOQSwab Lot Number: 2107946; Luminex Lot Number 080321023C, FLOQSwab Lot Number: 2107946; Luminex Lot Number 081621023D, FLOQSwab Lot Number: 2110850; Luminex Lot Number 081721023F, FLOQSwab Lot Number: 2107946, 2110850; Luminex Lot Number 081721023G, FLOQSwab Lot Number: 2110850; Luminex Lot Number 081921023D, FLOQSwab Lot Number: 2110850; Luminex Lot Number 082021023D, FLOQSwab Lot Number: 2100856; Luminex Lot Number 082021023E, FLOQSwab Lot Number: 2110850; Luminex Lot Number 082521023D, FLOQSwab Lot Number: 2100856, 2110850; Luminex Lot Number 082621023D, FLOQSwab Lot Number: 2110850; Luminex Lot Number 083121023D, FLOQSwab Lot Number: 2110850; Luminex Lot Number 090821023D, FLOQSwab Lot Number: 2100856; Luminex Lot Number 091521023D, FLOQSwab Lot Number: 2032870; Luminex Lot Number 091621023D, FLOQSwab Lot Number: 2100856; Luminex Lot Number 091621023G, FLOQSwab Lot Number: 2110850; Luminex Lot Number 092721023A, FLOQSwab Lot Number: 2110850; Luminex Lot Number 092821023E, FLOQSwab Lot Number: 2110850; Luminex Lot Number 092921023E, FLOQSwab Lot Number: 2110850; Luminex Lot Number 100621023C, FLOQSwab Lot Number: 2110850; Luminex Lot Number 101221023D, FLOQSwab Lot Number: 2110850; Luminex Lot Number 101321023D, FLOQSwab Lot Number: 2110850; Luminex Lot Number 101921023G, FLOQSwab Lot Number: 2110850; Luminex Lot Number 102021023E, FLOQSwab Lot Number: 2107946, 2110850; Luminex Lot Number 102221023D, FLOQSwab Lot Number: 2110850; Luminex Lot Number 102821023F, FLOQSwab Lot Number: 2107946.

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Kuwait, Austria, France, Greece, Kuwait, Turkey.