Dain Technology, Inc. (Subsidiary of Soundable Health, Inc.)
Dossier provenance
Source snapshots represented here
openFDA Device Enforcement
Report-date coverage
June 20, 2012–July 08, 2026
Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
openFDA Device Recall
event_date_initiated coverage
June 01, 1997–June 30, 2026
Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.
Product dates, classifications, firms, and source wording remain attached to their individual rows.
Opening this dossier never hides products that did not match a prior timeline filter.
device · product 1 of 1
ProudP Everyday Uroflow Tracker Software Version 1.x.x Software Version 1.2.0 or earlier
Due to interference with the Live Listen feature of hearing aid or AirPods, the user's iPhone may perform automatic processing of the urination sound signal, resulting in lower urination volume and velocity values than expected.
These labels are deterministic app interpretations, not FDA categories.
Due to interference with the Live Listen feature of hearing aid or AirPods, the user's iPhone may perform automatic processing of the urination sound signal, resulting in lower urination volume and velocity values than expected.
Code information
prouP Version - 1.2.0 UDI Code: (01)00860006184902(10)v1.x.x
Distribution pattern
Worldwide distribution - US Nationwide. The software app is distributed via the Apple App Store.