Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 90242

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 10, 2022
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
GE Healthcare, LLC

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

CARESCAPE Central Station (CSCS) V2, Model numbers 2082278-001 and 2082279-001. Update 3/10/2023: The following FRU and display part numbers have been added as affected devices: (1) Part #2082294-001, FRU MP200 POWER SUPPLY ASSEMBLY; (2) Part #2082293-002-R, FRU MP200 CPU ASSEMBLY; (3) Part #2082293-002, FRU MP200 CPU ASSEMBLY; (4) Part #2071523-001, MP200 22 IN DISPLAY WITH TOUCH; (5) Part #2092930-001, DSP LCD 22 IN MED W/TCH W/DKST N AMR; (6) Part #2092930-002, DSP LCD 22 IN MED W/TCH W/DKST BRIT P/C; (7) Part #2092930-003, DSP LCD 22IN MED W/TCH W/DKST EURO P/C; (8) Part #2092930-004, DSP LCD 22IN MED W/TCH W/DKST ITA P/C; (9), Part #2092930-005, DSP LCD 22IN MED W/TCH W/DKST INDIAN P/C; (10) Part #2092930-006, DSP LCD 22IN MED W/TCH W/DKST ISREAL P/C; (11) Part #2092930-007, DSP LCD 22IN MED W/TCH W/DKST SWISS P/C; (12) Part #2092930-008, DSP LCD 22IN MED W/TCH W/DKST AUSTRL P/C; (13) Part #2092930-009, DSP LCD 22IN MED W/TCH W/DKST JPN; (14) Part #2092930-010, DSP LCD 22IN MED W/TCH W/DKST BRZL P/C; (15) Part #2092930-011, DSP LCD 22IN MED W/TCH W DKST ARGNT P C; (16) Part #2092930-012, DSP LCD 22IN MED W TCH W DKST DENMARK; (17) Part #2092930-013, DSP LCD22IN MED W TCH W DKST S AFRICA P C; and (18) Part #2092930-014, DSP LCD22IN MED W TCH W DKST CHINA P C.

Z-1265-2022
Recall number
Z-1265-2022
Initiated
May 10, 2022
Classification
Class II
Status
Ongoing
Recalling firm
GE Healthcare, LLC
Quantity
30,993 Central Stations and 2,180 FRUs and displays

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
The CARESAPE Central Station (CSCS) V2 can shut down due to a potential power supply component failure. This can lead to loss of patient monitoring at the central station.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

The CARESAPE Central Station (CSCS) V2 can shut down due to a potential power supply component failure. This can lead to loss of patient monitoring at the central station.

Code information

All CARESCAPE Central Station V2 units: (1) CSCS V2 MA1700 Integrated, Product Code SKN, Model #2082278-001, GTIN 00840682109666; (2) CSCS V2 MAS700 Desktop, Product Code SNF, Model #2082279-001, GTIN 00840682109604. NOTE: The Product Code is the first 3 letters of the serial number.

Distribution pattern

Worldwide distribution - US Nationwide. There was also military/government distribution and the countries of Algeria, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Burkina Faso, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Cote D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, Germany, Ghana, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jersey, Kazakhstan, Kenya, Korea, Kuwait, Laos, Latvia, Lebanon, Libya, Lithuania, Macedonia, Madagascar, Malaysia, Maldives, Malta, Mauritius, Mexico, Mongolia, Montenegro, Morocco, Mozambique, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, North Macedonia, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Vietnam, and Yemen.