openFDA Device Enforcement
Report-date coverage
June 20, 2012–July 08, 2026
Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
openFDA Device Recall
event_date_initiated coverage
June 01, 1997–June 30, 2026
Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.
Product dates, classifications, firms, and source wording remain attached to their individual rows.
Opening this dossier never hides products that did not match a prior timeline filter.
device · product 1 of 2
Dimension LTNI -In vitro diagnostic test intended to quantitively measure cardiac troponin-I levels in human serum and heparinized plasma to aid in the diagnosis of myocardial infarction. Siemens Material Number (SMN)/REF (Catalog Number): 10444896/RF521
Positive bias with Lithium Heparin plasma samples from individuals that are expected to be below the 99th percentile of 0.07 ng/mL [0.07 ¿g/L], may lead to inappropriate intervention for myocardial infarction
These labels are deterministic app interpretations, not FDA categories.
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Positive bias with Lithium Heparin plasma samples from individuals that are expected to be below the 99th percentile of 0.07 ng/mL [0.07 ¿g/L], may lead to inappropriate intervention for myocardial infarction
Code information
UDI-DI: 00842768013904EA231420221110 Lot Number: EA2314 Exp. Date: 11-10-2022
Distribution pattern
Worldwide distribution - US Nationwide distribution in the states of AL, AR, CA, IL MN, MS, NC, ND, NH, OK, SD, TN, TX, WA and the countries of Argentina, Brazil, Canada, France, Germany, India, Mexico, Poland, Portugal, Slovakia, Spain.
device · product 2 of 2
Dimension CTNI-In vitro diagnostic test intended to quantitively measure cardiac troponin-I levels in human serum and heparinized plasma to aid in the diagnosis of myocardial infarction. Siemens Material Number (SMN)/REF (Catalog Number): 10444905/RF421C
Positive bias with Lithium Heparin plasma samples from individuals that are expected to be below the 99th percentile of 0.07 ng/mL [0.07 ¿g/L], may lead to inappropriate intervention for myocardial infarction
These labels are deterministic app interpretations, not FDA categories.
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Positive bias with Lithium Heparin plasma samples from individuals that are expected to be below the 99th percentile of 0.07 ng/mL [0.07 ¿g/L], may lead to inappropriate intervention for myocardial infarction
Code information
UDI-DI: 00842768013836GA232320221119 Lot Number: GA2323 Exp. Date: 11-19-2022 Recall Extended 9/21/22: UDI: 00842768013836BA308820230329 Lot Number: BA3088 Exp. Date: 2023-03-29
Distribution pattern
Worldwide distribution - US Nationwide distribution in the states of AL, AR, CA, IL MN, MS, NC, ND, NH, OK, SD, TN, TX, WA and the countries of Argentina, Brazil, Canada, France, Germany, India, Mexico, Poland, Portugal, Slovakia, Spain.