Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 90387

2 recalled-product records grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
May 18, 2022
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Siemens Healthcare Diagnostics, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

2 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 2

Dimension LTNI -In vitro diagnostic test intended to quantitively measure cardiac troponin-I levels in human serum and heparinized plasma to aid in the diagnosis of myocardial infarction. Siemens Material Number (SMN)/REF (Catalog Number): 10444896/RF521

Z-1340-2022
Recall number
Z-1340-2022
Initiated
May 18, 2022
Classification
Class II
Status
Terminated
Quantity
76 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Positive bias with Lithium Heparin plasma samples from individuals that are expected to be below the 99th percentile of 0.07 ng/mL [0.07 ¿g/L], may lead to inappropriate intervention for myocardial infarction

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Positive bias with Lithium Heparin plasma samples from individuals that are expected to be below the 99th percentile of 0.07 ng/mL [0.07 ¿g/L], may lead to inappropriate intervention for myocardial infarction

Code information

UDI-DI: 00842768013904EA231420221110 Lot Number: EA2314 Exp. Date: 11-10-2022

Distribution pattern

Worldwide distribution - US Nationwide distribution in the states of AL, AR, CA, IL MN, MS, NC, ND, NH, OK, SD, TN, TX, WA and the countries of Argentina, Brazil, Canada, France, Germany, India, Mexico, Poland, Portugal, Slovakia, Spain.

device · product 2 of 2

Dimension CTNI-In vitro diagnostic test intended to quantitively measure cardiac troponin-I levels in human serum and heparinized plasma to aid in the diagnosis of myocardial infarction. Siemens Material Number (SMN)/REF (Catalog Number): 10444905/RF421C

Z-1341-2022
Recall number
Z-1341-2022
Initiated
May 18, 2022
Classification
Class II
Status
Terminated
Quantity
113 units; Extended: 52 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Positive bias with Lithium Heparin plasma samples from individuals that are expected to be below the 99th percentile of 0.07 ng/mL [0.07 ¿g/L], may lead to inappropriate intervention for myocardial infarction

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Unknown

An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Positive bias with Lithium Heparin plasma samples from individuals that are expected to be below the 99th percentile of 0.07 ng/mL [0.07 ¿g/L], may lead to inappropriate intervention for myocardial infarction

Code information

UDI-DI: 00842768013836GA232320221119 Lot Number: GA2323 Exp. Date: 11-19-2022 Recall Extended 9/21/22: UDI: 00842768013836BA308820230329 Lot Number: BA3088 Exp. Date: 2023-03-29

Distribution pattern

Worldwide distribution - US Nationwide distribution in the states of AL, AR, CA, IL MN, MS, NC, ND, NH, OK, SD, TN, TX, WA and the countries of Argentina, Brazil, Canada, France, Germany, India, Mexico, Poland, Portugal, Slovakia, Spain.