openFDA Device Enforcement
Report-date coverage
June 20, 2012–July 08, 2026
Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
openFDA Device Recall
event_date_initiated coverage
June 01, 1997–June 30, 2026
Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.
Product dates, classifications, firms, and source wording remain attached to their individual rows.
Opening this dossier never hides products that did not match a prior timeline filter.
Several complaints were received for the affected lot; users received a "Fiber type does not match card type" error message on the GreenLight XPS Laser console, necessitating the replacement of the laser fiber and fiber card. This may result in prolonged procedure while the fiber is being replaced.
These labels are deterministic app interpretations, not FDA categories.
Several complaints were received for the affected lot; users received a "Fiber type does not match card type" error message on the GreenLight XPS Laser console, necessitating the replacement of the laser fiber and fiber card. This may result in prolonged procedure while the fiber is being replaced.
Code information
UDI-DI (GTIN): 00878953005522 Lot: 27770466
Distribution pattern
Domestic distribution to AR CA IL ME MI NC TX Foreign distribution to Australia, Canada, and France.