Description/Part: Simplify Disc Size 1, Ht 5/SM-5; Simplify Disc Size 1, Ht 6/SM-6; Simplify Disc Size 2, Ht 4/MD-4; Simplify Disc Size 2, Ht 5/MD-5; Simplify Disc Size 2, Ht 5, 5/MD-5L; Simplify Disc Size 2, Ht 6/MD-6; Simplify Disc Size 2, Ht 6, 5/MD-6L; Simplify Disc Size 3, Ht 5/LG-5; Simplify Disc Size 3, Ht 5, 5/LG-5L; Simplify Disc Size 3, Ht 6/LG-6; Simplify Disc Size 3, Ht 6, 5/LG-6L
- Recall number
- Z-1453-2022
- Initiated
- June 20, 2022
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- NuVasive Inc
- Quantity
- 1,963
App-derived interpretation
Cervical artificial disc Unique Device Identifier (UDI) barcodes contain the incorrect Global Trade Item Number (GTIN) on the Patient Label when scanned and may also contain the incorrect GTIN human readable text, which could lead to selecting an incorrectly sized unit.
Official device-enrichment evidence · Sourced
Labeling Change Control
Inspect official wording and provenance
Reason for recall
Cervical artificial disc Unique Device Identifier (UDI) barcodes contain the incorrect Global Trade Item Number (GTIN) on the Patient Label when scanned and may also contain the incorrect GTIN human readable text, which could lead to selecting an incorrectly sized unit.
Code information
Part/Correct GTIN (DI)/Lot: SM-5/843285100023/W79256, W82070, W80144; SM-6/843285100047/W80703, W81624, W80701, W81132, W81133, W81633, W81625, W81611, W81828, W81971, W82544, W82543, W81034, W81975, W81580, W81612, W81621, W81131, W81613, W80909, W81623, W81634, W79466, W81033, W81134, W81600, W79259; MD-5/843285100061/W81136, W81135, W80704, W80700; MD-5L/843285100078/W80707, W80706; MD-6/843285100085/W80911, W81642, W81643, W82545, W82546; MD-6L/843285100092/W81822; LG-5/843285100108/W81138, W81141, W79465, W81994, W81142, W81626, W81143, W79257; LG-5L/843285100115/W79258, W80146, W80915, W82548; LG-6/843285100122/W82547; LG-6L/843285100139/W81827
Distribution pattern
US: CA, WA, TX, IN, MO, MI, FL, VA, OR. OUS: DE, NZ, SG, UK