Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards and may decrease the overall reliability of the system and result in delay of treatment
These labels are deterministic app interpretations, not FDA categories.
Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards and may decrease the overall reliability of the system and result in delay of treatment
Code information
UDI-DI: 00821925044197 Serial numbers: CG5085 and CG6009
Distribution pattern
US Nationwide distribution in the states of AL, NM.