Recall events
/
Event 90629
Event summary
Timeline bucket July 15, 2022
Product types Device
Classifications Class II
Statuses Terminated
Recalling firm wording DeRoyal Industries Inc
Dossier provenance
Source snapshots represented here
openFDA Device Enforcement
Report-date coverage
June 20, 2012–July 08, 2026
Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
openFDA Device Recall
event_date_initiated coverage
June 01, 1997–June 30, 2026
Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.
How coverage and grouping work
Complete imported group
Every recalled product in this event
1 official enforcement rows
Product dates, classifications, firms, and source wording remain attached to their individual rows.
Opening this dossier never hides products that did not match a prior timeline filter.
device · product 1 of 1
DeRoyal limb holders: (1) DeRoyal Limb Holder, Wrist, Fixed, Quick Release, REF M8124; (2) DeRoyal Limb Holder, Ankle, Fixed, Quick Release, REF M8125; (3) DeRoyal Security Cuffs, Wrist and Ankle, Universal, REF M8136; and (4) DeRoyal Security Cuffs, Heavy Duty Vertical Psychiatric, University, REF M8137.
Z-1562-2022
Recall number Z-1562-2022
Initiated July 15, 2022
Classification Class II
Status Terminated
Quantity 36 eaches and 7 pairs
App-derived interpretation
Unknown
reason.no_named_rule · v1.0.0
The webbing strap was laced improperly through the teeth of the buckle allowing the webbing strap to slide and was, therefore, not secure.
These labels are deterministic app interpretations, not FDA categories.
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Enrichment source openFDA Device Recall
Exact join Official recall number Z-1562-2022
Source locator 0001-device-recall-0001-of-0001.json.zip#results[33634]
Snapshot SHA-256 77e96a0ce750de9fa4bb4737d40b653784372b5abe88b9de499ad51edd808a4b
Open official device record
Compare all device evidence states
Inspect official wording and provenance
Reason for recall The webbing strap was laced improperly through the teeth of the buckle allowing the webbing strap to slide and was, therefore, not secure.
Code information (1) M8124 - Lot 57372551 / exp 06/06/2032; Lot 57285721 / exp 05/22/2032; Lot 57369651 / exp 06/06/2032; Lot 57497652 / exp 6/27/2032; GTIN: 00749756780728; (2) M8125 - Lot 5746744 / exp 6/20/2032; Lot 57419195 / exp 6/13/2032; Lot 57420031 / exp 6/14/2032; GTIN: 00749756780735; (3) M8136 - Lot 57501121 / exp 6/27/2032; GTIN: 00749756837828; (4) M8137 - Lot 57462151 / exp 6/20/2032; GTIN: 00749756849234.
Distribution pattern US Nationwide distribution in the states of FL, IN, NV, RI, TN, TX, and WV.
Official fact provenance
Dataset openFDA Device Enforcement
Snapshot published July 16, 2026 07:03
Snapshot SHA-256 1c273507f2e0fc8c5b330c3716a1392933bd90a42a460afcb7368f298d5cbd3a
Raw source locator 0001-device-enforcement-0001-of-0001.json.zip#results[37592]
FDA event record
· Exact recall-number query on openFDA