Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 90777

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
August 01, 2022
Product types
Device
Classifications
Class II
Statuses
Terminated
Recalling firm wording
Randox Laboratories Ltd.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

The Randox Serology I Positive Control contains the following analytes in human based plasma: Anti HIV 1/2, Anti HTLV 1/2, Anti HCV, HBsAg, Anti HBc, and Anti Treponema pallidum IgG. The product is delivered as a liquid in a kit of 3 identical ready-to-use vials of 5 mL each.

Z-1772-2022
Recall number
Z-1772-2022
Initiated
August 01, 2022
Classification
Class II
Status
Terminated
Recalling firm
Randox Laboratories Ltd.
Quantity
2 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
***Update** :HBsAg analyte is also Non-Reactive on the Ortho Vitros analyser as well as the Beckman Coulter DxI. Randox Serology I Positive Control will test negative (Non-Reactive) for Marker HBsAg resulting in delay in reporting patient results due to the positive control O36SR testing negative (Non-Reactive) for Marker HBsAg

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process change control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

***Update** :HBsAg analyte is also Non-Reactive on the Ortho Vitros analyser as well as the Beckman Coulter DxI. Randox Serology I Positive Control will test negative (Non-Reactive) for Marker HBsAg resulting in delay in reporting patient results due to the positive control O36SR testing negative (Non-Reactive) for Marker HBsAg

Code information

GTIN: 05055273216509 Lot number: 036SR Exp Date: 28th March 2023

Distribution pattern

US Nationwide distribution in the states of WV, NC.