device · product 1 of 1
IMMULITE 2000 and IMMULITE 2000 XPi Thyroid Stimulating Immunoglobulins (TSI) Siemens Material Number (SMN): 10876421 US; 10713448 (OUS)
- Recall number
- Z-1820-2022
- Initiated
- August 11, 2022
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 100 units US; 328 units OUS
App-derived interpretation
An average negative bias of -23% with IMMULITE 2000/IMMULITE 2000 XPi Thyroid Stimulating Immunoglobulins (TSI) when compared to other in-date kit lots, there is a potential for erroneously depressed patient results. This may lead to a delayed follow up of patients with clinical autoimmune thyroid
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
Inspect official wording and provenance
Reason for recall
An average negative bias of -23% with IMMULITE 2000/IMMULITE 2000 XPi Thyroid Stimulating Immunoglobulins (TSI) when compared to other in-date kit lots, there is a potential for erroneously depressed patient results. This may lead to a delayed follow up of patients with clinical autoimmune thyroid
Code information
UDI-DI: (01)00630414597188(10)D389S(17)221130 (US); (01)00630414597171(10)389(17)221130 (OUS) Lot Number: 389 Exp.Date: 2022-11-30
Distribution pattern
Worldwide Distribution: US (nationwide) to states of: FL IN MA MI NH NJ NY OH TX UT and OUS (Foreign) countries of: Australia¿¿¿¿¿¿ Austria¿¿¿¿¿¿¿¿ Brazil¿¿¿¿¿¿¿¿¿ Czech¿Republic¿ Germany¿¿¿¿¿¿¿¿ Italy¿¿¿¿¿¿¿¿¿¿ P.R.¿China¿¿¿¿¿ Singapore¿¿¿¿¿¿ Spain¿¿¿¿¿¿¿¿¿¿ Taiwan¿¿¿¿¿¿¿¿¿ Tunisia¿¿¿¿¿¿¿¿ Turkey¿¿¿¿¿¿¿¿¿