device · product 1 of 1
Ambra PACS, Ambra ProViewer. Software for use as a primary diagnostic and analysis tool for diagnostic images.
- Recall number
- Z-1818-2022
- Initiated
- June 23, 2022
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- DICOM Grid, Inc. d/b/a Ambra Health
- Quantity
- 671 downloads
App-derived interpretation
A race condition between the storage system and services database has the potential to revert edits made to patient information upon first ingestion of a study.
Official device-enrichment evidence · Sourced
Software design
Inspect official wording and provenance
Reason for recall
A race condition between the storage system and services database has the potential to revert edits made to patient information upon first ingestion of a study.
Code information
UDI-DI: AMBRHEALTHSOLUTIONS0; Software Version: 3.22.2.0 and 3.22.3.0.
Distribution pattern
Domestic: AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, NE, NV, NH, NJ, NM, NY, NC, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, DC, & PR.