Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 90839

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
August 31, 2022
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Micro-X Ltd.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

MICRO-X ROVER MOBILE X-RAY SYSTEM, MXU-RV19

Z-1786-2022
Recall number
Z-1786-2022
Initiated
August 31, 2022
Classification
Class II
Status
Ongoing
Recalling firm
Micro-X Ltd.
Quantity
42 units

App-derived interpretation

Device software or design reason.device_software_design · v1.0.0
Software calibration error

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process control

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Software calibration error with product equip with a Dose Area Product (DAP) meter. This results in the product displaying the incorrect DAP meter values which may impede radiation exposure management decisions

Code information

Model Number: MXU-RV19 UDI Codes: (01)09357123000013(11)210326(21)00222 (01)09357123000013(11)210628(21)00234 (01)09357123000013(11)210629(21)00235 (01)09357123000013(11)210629(21)00236 (01)09357123000013(11)210630(21)00237 (01)09357123000013(11)210708(21)00238 (01)09357123000013(11)210721(21)00240 (01)09357123000013(11)210910(21)00261 (01)09357123000013(11)210913(21)00262 (01)09357123000013(11)210914(21)00263 (01)09357123000013(11)210921(21)00264 (01)09357123000013(11)210922(21)00265 (01)09357123000013(11)210923(21)00266 (01)09357123000013(11)210927(21)00267 (01)09357123000013(11)211006(21)00268 (01)09357123000013(11)211008(21)00269 (01)09357123000013(11)211011(21)00270 (01)09357123000013(11)211011(21)00271 (01)09357123000013(11)211012(21)00272 (01)09357123000013(11)211013(21)00273 (01)09357123000013(11)211013(21)00274 (01)09357123000013(11)211210(21)00288 (01)09357123000013(11)211213(21)00289 (01)09357123000013(11)211213(21)00290 (01)09357123000013(11)220329(21)00305 (01)09357123000013(11)220330(21)00306 (01)09357123000013(11)220331(21)00307 (01)09357123000013(11)220401(21)00308 (01)09357123000013(11)220404(21)00309 (01)09357123000013(11)220405(21)00310 (01)09357123000013(11)220406(21)00311 (01)09357123000013(11)220404(21)00312 Serial Numbers: 222 234 235 236 237 238 240 261 262 263 264 265 266 267 268 269 270 271 272 273 274 288 289 290 305 306 307 308 309 310 311 312

Distribution pattern

Worldwide distribution - U.S. Nationwide distribution in the states of CA, NV, NY, OH, TX and WA and the countries of Australia, France, Israel, Japan, and United Arab Emirates.