WEB Detachment Controller, REF: WDC-2, WDC, MV-WDC2. Component of WEB Aneurysm Embolization System.
- Recall number
- Z-0170-2023
- Initiated
- September 21, 2022
- Classification
- Class II
- Status
- Ongoing
- Recalling firm
- MICROVENTION INC.
- Quantity
- 2018
App-derived interpretation
out of specification
Official device-enrichment evidence · Sourced
Process control
Inspect official wording and provenance
Reason for recall
Detachment controller, of an aneurysm embolization system, has an out of specification tunnel opening dimension, which may prevent the implant device pusher from being fully inserted into the controller, which may result in the inability of the controller to properly detach the implant device, which could cause delays and excessive manipulation that may cause the implant to move and damage vessels
Code information
REF/UDI/Lots: WDC-2/00842429103128/0000095683, 0000096284, 0000205967, 0000206629, 0000215962, 0000216862, 0000224074, 0000241662, 0000242343, 0000242345; Distributed in China: WDC/00842429113684/0000078840, 0000086571, 0000088347, 0000088349, 0000136477, 0000181460, 0000191147; Distribution in Japan: MV-WDC2/04987892122231/0000142172; 0000166445; 0000210109; 0000245337
Distribution pattern
Worldwide - US Nationwide distribution in the states of MN, PA, NY, MI, FL, VA, TX, OH, CA, NE, MN, NC, LA, IL, TN, IN, UT, KS and the countries of Japan, China, Argentina, Australia, South Korea, Italy, UK.